98
FUNCTION SETTING USING SETTINGS SCREEN
LIST OF SETUP ITEMS
Setup items are formed with eight indexes for each subject.
"MEASURE" Set items related to measurement functions
"INITIAL"
Set items related to the initial status after power on
"DISPLAY"
Set items related to screen display
"PRINT INT" Set items related to output from the internal printer
"PRINT EXT" Set items related to output from the external printer
"COMM"
Set items related to data input/output between the external device
"NETWORK" Set items related to input/output using the LAN
"SPECIAL"
Set items related to maintenance (for service engineer only)
MEASURE (TEMPORARY SETTING)
MEASURE contains settings related to temporarily changeable measurement functions.
Unlike initial setting (INITIAL), this function is used to temporarily perform measurements
under certain conditions.
After power on, clearing all measurement values, print-out and data output, the system
returns to the initial status.
Set item display
Setting Change button
Set contents
Initial value
Page
R-L Measure
BINO
Set to both-eye measurement
BINO
MONO
Set to single-eye measurement
Auto Mode
FULL AUTO
Set to (FULL) AUTO mode
FULL AUTO
AUTO TRACKING
Set to AUTO TRACKING mode
MANUAL
Set to MANUAL mode
Cont. Cycle
1
Sets the number of continuous measurements under
AUTO mode, using up/down buttons.
3
2
3
4
5
6
7
8
9
10
IOL Mode
OFF
IOL mode is not set.
OFF
ON (300ms)
IOL mode is set, with exposure time of 300ms.
ON (400ms)
IOL mode is set, with exposure time of 400ms.
ON (500ms)
IOL mode is set, with exposure time of 500ms.
Target Brightness
Level 1
Target brightness is set to the darkest of 4 levels.
Level 3
Level 2
Target brightness is set to the 2nd darkest of 4 levels.
Level 3
Target brightness is set to the 3rd darkest of 4 levels.
Level 4
Target brightness is set to the brightest of 4 levels.
Cyl sign
+
Display style of astigma sign is set to "+".
-
-
Display style of astigma sign is set to "-".
MIX
Display style of astigma sign is set to "MIX".
Summary of Contents for KR-1W
Page 1: ...INSTRUCTION MANUAL WAVE FRONT ANALYZER KR 1W ...
Page 2: ......
Page 186: ......
Page 188: ...KR 1W WAVE FRONT ANALYZER 41835 97992 Printed in Japan 2010 02 100LW2 ...