122
FUNCTION SETTING USING SETTINGS SCREEN
PACKING MODE
Set the instrument (KR-1W) to the packing position (as-delivered condition).
1
Tap the Set Condition button of the set item display "Packing Mode" of "INITIAL."
2
A message "Run the Packing mode?" is displayed.
3
When "Yes" is selected, the operation starts automatically. A message "Wait until the
packing operation is completed." will be displayed.
4
When the operation is complete, a message "Packing mode has been completed. Shut off
the power directly." will be displayed.
5
Setting is completed by shutting off the power.
PUPIL DISTANCE
The travel distance of both eyes can be set from 58mm to 74mm. Before shipment, the default
setting is "68mm."
1
Tap the Set Condition button of the set item display "Pupil Distance" of "INITIAL."
2
Enter the number by the ten-key display of the Setting Change button .
3
Setting is done by tapping the OK button .
When moving from the measuring position of one eye to the other eye, the total R-
L travel distance set does not mean the device will travel this entire distance; the
actual travel distance is 1/2 of the set distance from the origin/center of R-L move-
ment for both right and left. Thus, if the setting is "68mm" as shipped, the travel
distance is 34mm from the center.