MESSAGE LIST
START
"Paper end. Please set the
paper."
Paper end. Please set new paper.
"Close printer cover."
Printer cover is open.
MEASUREMENT SCREEN
"OVER-SPH"
Spherical power e22D or -25D (at VD=12mm).
"OVER-CYL"
Cylindrical power exceeds ±10D (at VD=12mm).
"OVER-R"
Corneal curvature exceeds 5.00-10.00mm.
"NO TARGET"
No patient's eye or the eye image is too dark.
"NO CENTER"
Not possible to find the center of patient's eye.
"Please move to center eye,
and focus on."
The eye center is off the alignment range. Touch the eye center
on the control panel.
"Please focus on eye"
Tilt the control lever back and forth so as the ring look clearer.
"No measurement data and
analyzed result."
No measurement was done. Perform measurements.
"Close printer cover. Please
push OK button and DATA
OUT button after you close
the cover."
Printer button is pushed with the printer cover open.
"Paper end. Please push
OK button and DATA OUT
button after you set the
paper."
Printer paper is used up.
"Unable to connect the
network."
Displayed when the KR-1W data output is turned ON and the
DATA OUT button is pressed, without connecting the PC.
"Failed to send the data via
network".
Ethernet data transmission has failed.
"Nothing is output in this
setting. Please change the
settings."
When settings are as follows, this is displayed by tapping DATA
OUT button:
• Initial setting:
Thermal printer (Do not use)
• Data communication: RS-232C output setting (OFF)
• LAN connection:
KR-1W data output (OFF)
• USB:
Not connected
"Outputting data..."
Data exporting...
"Data output finished."
Data export has been completed.
"ERROR : Failed to send
the data via RS-232C"
RS232C data transmission has failed.
"ERROR : Failed to send
the data via USB"
USB data transmission has failed.