SPECIFICATIONS
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC Standard (IEC 60601-1-2 Ed.3.0:2007).
Item
Parts code
Cable shield
Ferrite core
Length (m)
AC power cord
4480470020
No
No
1.5
RS-232C CROSSING CABLE
418120002
Yes
Yes
3.0
USB CABLE
-
No
No
3.0
LAN CABLE
-
No
Yes
3.0
Guidance and manufacturer's declaration - electromagnetic emissions
The KR-1W is intended for use in the electromagnetic environment specified below.
The customer or the user of the KR-1W should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The KR-1W uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The KR-1W is suitable for use in all establish-
ments other than domestic and those directly con-
nected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided in the ACCOMPA-
NYING DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM
as replacement parts for internal components, may result in increased EMISSIONS or
decreased IMMUNITY of the EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment.
IF adjacent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to
verify normal operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than
those specified may result in increased EMISSION or decreased IMMUNITY of the EQUIP-
MENT or SYSTEM.