The chinrest moves up/down while the
button or
button is pressed.
The chinrest stops if the applied load exceeds the specified weight. If the chinrest
does not work even after removing the chin from it, it may be at fault. Turn OFF the
POWER switch and disconnect the power cable: Call your dealer or TOPCON
at the address printed on the back cover of this manual.
If the applied load exceeds the specified weight, the chinrest may not work. Advise
the patient not to push down on the chinrest more than necessary.
If the patient's head is placed in the center of the chinrest, measurement may be
disabled with a limit message. Make sure the patient's head is placed correctly.
For details about the limit message, see "Limit message" on page 40.
Adjust the height of the automatic instrument table so that the patient
can sit on the chair comfortably and correct measurement values can be
obtained.
Measurement in auto mode may not be possible due to frequent blinks
or existing abnormalities in the corneal surface caused corneal disease
etc.
In this case, use manual mode.
When operating the instrument, be careful not to touch the patient's face
or nose. If touched, clean the instrument following "CLEANING THE
INSTRUMENT" on page 154.
39
BASIC OPERATIONS
MEASUREMENT UNDER AUTO MODE
SETTING THE AUTO MODE
1
Confirm the waiting status for measurement.
Confirm that the AUTO button of the control panel is displayed in orange. If not, tap the
AUTO button, select Auto mode (Full Auto), and set the waiting status for measure-
ment.
Alignment mark
Check eye level and forehead,
press measure switch.
Luminous point
AUTO button
Pupil image
Ring image
2
A message "Check eye level and forehead, press measure switch." is displayed.
Confirm that the pupil image and luminous point are displayed on the screen.
MEMO
MEMO
MEMO