2.Measurement Operation
25
2.1.3 Absolute Value Measurement
The following describes the procedure for measuring absolute values.
1
Set to the absolute value measurement mode (ABSOLUTE) in the function mode.
☞
"3.1.1 Entering/Returning from the Function Mode"
2
Press the [RUN/HOLD] switch to start measurement.
3
To stop measurement, press the [RUN/HOLD] switch.
2.1.4 Measuring Difference
On this instrument, you can measure the difference from reference data.
The following explains the procedure for measuring difference.
1
Set to the difference measurement mode (DIFFERENCE) in the function mode.
☞
"3.1.1 Entering/Returning from the Function Mode"
2
The following screen is displayed when the difference mode is set after the instrument is
turned ON.
3
Set the reference sample in place, and press the [RUN/HOLD] switch.
After measurement, the measurement result is displayed for about 3 seconds, then
continuous difference measurement is performed automatically.
Request
・
Reference data is stored internally until the power is turned OFF. The reference
data is not measured even if DIFFERENCE is set after DIFFERENCE is
changed to ABSOLUTE in the function mode. To change the reference data,
the power OFF then back ON again, and repeat the procedure from step 1.
・
Stored reference data can be checked in the function mode.
Checking reference data
☞
"3.2.2.1 Measuring Difference"
Memo