2.Measurement Operation
35
■ Analog output response speed
The analog output response speed differs depending on the measurement range used.
Response speed
☞
"Appendix: Analog Output Response Speed"
To observe the characteristics of a flashing light source, set the analog output response speed in the
function mode to FAST. When the speed is set to NORMAL, the integral capacitor value is increased
and operation undergoes smoothing, so the response speed becomes slow.
Function mode
☞
"3.2.11 Selecting the Analog Output Response Speed"
■How to use analog output
The following describes the procedure for observing analog output.
1
Solder and attach the shielded wire of the core to the analog output plug (supplied), and
connect it to the device.
・
The pin side of the plug is the signal.
2
Set the measurement range, measurement method, tristimulus value filter, and analog output
response speed in the function mode.
Function mode
☞
"3.2.8 Setting the Measurement Range"
Function mode
☞
"3.2.9 Setting the Measurement Mode"
Function mode
☞
"3.2.10 Selecting the Filter of the Tristimulus Value Fixed Mode"
Function mode
☞
"3.2.11 Selecting the Analog Output Response Speed"
3
Press the [RUN/HOLD] switch to start measurement.
・
Analog output can also be observed in a [HOLD] state.
Request
・
Correct measurement cannot be performed when this feature is used with
output voltage in a saturated state. Be sure to set the proper
measurement range.
・
The output impedance is about 100Ω. Use a recorder having an input
impedance
of 10kΩ or more.
Memo
Memo