3. Various Settings
47
Measurement results display
4
The following function mode screen is displayed. Press the [FUNCTION] switch to return to
the measurement mode.
5
Press the [RUN/HOLD] switch to start measurement.
3.2.3 Setting the Averaging Count
Averaging measurement is performed by selecting [AVE.(AVERAGE)] on the [SINGLE/AVE.] panel
switch. In this section, select the number of averaging measurements.
The averaging count can be selected from 2 times/3 times/5 times/10 times.
Display method
☞
"3.1.2 Displaying Data/Setting Items"
Selecting the [SINGLE/AVE.] panel switch
☞
"1.2 Part Names and Functions"
1
Press the [CHANGE] button.
2
Press the [ROTATION] button and select the setting.
Each press switches the display as follows:
2 => 3 => 5 => 10 => 2
…..
3
After you have selected the setting, press the [ENTER] button.
・
The absolute value measurement/difference measurement setting is held in memory even
if the instrument is turned OFF.
・
The standard sample data is cleared from memory by turning the power OFF then back
ON again.
・
To check the standard sample stored to memory, press the [ENTER] switch in the
screen in Step 1.
Memo
*
W
S T A N D A R D
t
S A M P L E
T 0
*
P l e a s e
x =
e
0 . 3 1 2 7
3 + 0 0 *
t h e Y S M
y =
s
0 . 3 2 9 0
t h e n *
p
L =
h
1 . 0 0 0 E + 0 2
0
c d / m ^ 2
F 2 . 0
*
D I F
R
K 0 0
A
G 0 0 - 0
0 *
P l e a s e
x =
e
0 . 0 1 8 3
3 + 0 0 *
t h e Y S M
y =
s
0 . 0 2 2 7
t h e n *
p
L = - 1 . 1 2 1 E + 0 1
0
c d / m ^ 2
*
2
A V E R A G E
K
M E A S U R E
0 0
*
P l e a s e x = e 0 . 0 1 8 3 3 + 0 0 *
t h e Y S
A V E R A G E
2
5
7 t h e n *
p L = - 1 . 1 2 1 E + 0 1 0 c d / m ^ 2