4. Communication With a PC
80
4.2.17 FKR Command
Reads the current correction type.
1: Normal correction 2: Direct correction
PC
Command
BM-5AC
FKR+(Cr
・
Lf)
OK+(Cr
・
Lf)
Correction type
END+(Cr
・
Lf)
4.2.18 FAG Command
Enables area correction of the instrument.
FAG [n]
:
Number of the area correction factor group to be enabled. 1 to 10
PC
Command
BM-5AC
FAG2+(Cr
・
Lf)
OK+(Cr
・
Lf)
4.2.19 FO Command
(O: alphabet)
Disables area correction of the instrument.
PC
Command
BM-5AC
FO+(Cr
・
Lf)
OK+(Cr
・
Lf)
4.2.20 CGL Command
Clears the chromaticity area in the area correction factor per each individual group.
CGL [n]
:
Numberof area correction factor group whose chromaticity area is to be
cleared. 1 to 10
PC
Command
BM-5AC
CGL1+(Cr
・
Lf)
OK+(Cr
・