4. Communication With a PC
84
4.3 Remote Measurement Output Format
The output format of the data from the instrument during remote measurement (ST command) is
as follows.
■BM-5AC mode
Row
Number
Output Data
Example
Data Content
1
D*
Indicates the brightness of the measurement target with respect to
the measurement range of the instrument.
D0: Normal D1: Under D2: Over
Under range-D1, all of tristimulus values X
2
Y Z are displayed in
set range below the value of the following table.
Measuring
Field
Tristimulus value (range 1)
The values in ranges
2 to 5 are values
obtained by
multiplying 10X,
100X, 1000X, and
1000X, respectively.
X
2
Y
Z
3
゚
2
゚
1
゚
0.2
゚
0.1
゚
0.008
0.018
0.072
1.8
7.2
0.009
0.020
0.080
2.0
8.0
0.009
0.020
0.080
2.0
8.0
2
M*
Measurement mode
M0: xyL mode M1: u'v'L mode M2: Tc/duv/L mode
3
T*
Measurement averaging
TF: Single measurement TS: Averaging measurement
4
R*
Measurement range setting
RA0: AUTO range (common range)
RA1: AUTO range (individual range)
RM0: Manual range (common range)
RM1: Manual range (individual range)
5
X*
X2 Range (range X1 to X5)
6
Y*
Y range (range Y1 to Y5)
7
Z*
Z range (range Z1 to Z5)
8
UC
Unit UC: cd/
㎡
9
F*
Measuring Field
F1: 0.1° F2: 0.2° F3: 1° F4: 2° F5: 3°
10
K*
Correction factor K0: No correction K1 to K9: Correction factor
number.
11
FG*
Area correction group number.
FG0: No correction
FG1 to FG10: Area correction group number
12
GK*
Applicable chromaticity area
GK0: No applicable chromaticity area
GK1 to GK5: Color data number corresponding to chromaticity
area
13
1.234E+56
Luminance
14
1.234E+56
Tristimulus value X
15
1.234E+56
Tristimulus value Y
16
1.234E+56
Tristimulus value Z
17
1.234
Chromaticity x
18
1.234
Chromaticity y
19
1.234
Chromaticity u'
20
1.234
Chromaticity v'
21
12345
Color temperature
22
123456
Deviation
23
END
Data end command