4. Communication With a PC
86
■BM-5A mode communication format
Output data are measurement conditions and various measurement color values.
There are three output formats as follows depending on the measurement mode.
・
M0 command-
DnT*R*RnU*Fn_
X=_#.#### _ y=_#.#### _ X=_#.###E±## _ Y=_#.###E±## _ Z=_#.###E±##CR
・
M1 command-
DnT*R*RnU*Fn_
u'=_#.#### _ v'=_#.#### _ X=_#.###E±## _ Y=_#.###E±## _ Z=_#.###E±##CR
・
M2 command-
DnT*R*RnU*Fn_
Tc=_##### _ duv=_#.#### _ X=_#.###E±## _ Y=_#.###E±## _ Z=_#.###E±##CR
○Explanation of measurement conditions
・
Dn: Judgment of measurement status
n=0 ... Normal
1 ... Over range: This is displayed when at least 1 of tristimulus values X
2
Y Z exceeds
the setting range.
2 ... Under range: This is displayed when all of tristimulus values X
2
Y Z are at the
values in the table below or lower.
Values for which is "D2" displayed
Measuring
Field
Tristimulus value (range 1)
X
2
Y
Z
3
゚
0.008
0.009
0.009
2
゚
0.018
0.020
0.020
1
゚
0.072
0.080
0.080
0.2
゚
1.8
2.0
2.0
0.1
゚
7.2
8.0
8.0
The value of the above table is a design value.
It is likely to differ somewhat in each product.
・
T* : Measurement count
TF : 1 measurement,
TS : Average of 5 measurements
・
R* : Measurement range
RA : AUTO,
RM : MANUAL
・
Rn : Measurement range used for measurement
n
: 1 to 5
・
U* : unit
UC : cd/m
2
・
Fn : Measuring field
F1 : 0.1°, F2: 0.2°, F3: 1°, F4: 2°, F5: 3°
* _ indicates a space.
* # indicates a numerical value.
* The carriage return code (0x0d) is indicated by CR.
* When the CR(0x0d) carriage return code is not read by the computer, this instrument judges that
transmission is incomplete, and the instrument enters a non-operable state.
※
The values in ranges 2 to 5 are
values obtained by multiplying
10X,
100X,
1000X,
and
1000X, respectively.