58
E0007
E0009
Failure of RS-232C initialization
Failure of LAN initialization
Inner device might be damaged
Restart the instrument. If the instrument does not recover from the
error, Please contact your dealer or Topcon Technohouse. Advice us of
the error code, the circumstance when the problem happen in details.
If momentary interruption of power, let us know it also.
RS-232C init error
LAN init error
E0008
E0010
Error happen in RS-232C communication
Error happen LAN communication
Check following points;
-
Communication setting in PC and the instrument
-
Cable and connection in PC and the instrument
RS-232C trans error
LAN trans error
E0011
E0012
Under range error happen
Over range error happen
Check to see if the luminance of the object is within the measurable
range.
under range error
over range error
E0014
Entry data is beyond valid value range.
Enter the data within valid value range
value out of range
E0019
E0020
E0021
E0022
E0023
E0024
・
・
・
E0031
E0032
E0033
E0034
E0035
E0036
Error happen in motor or sensor
Error or sensory might be damaged
Restart the instrument. If the instrument does not recover from the
error, Please contact your dealer or Topcon Technohouse. Advice us of
the error code, the circumstance when the problem happen in details.
If momentary interruption of power, let us know it also.
motor* init error
motor* origin error
motor* move error
motor* sensor error
motor* signal error
motor* pos error