45
3
“CALCULATING” will flash indicating that the
calibration value is being calculated by the
instrument. Do not touch the instrument until “END”
is displayed to signify operation completed. (If you
touch the instrument, you will need to recalibrate.)
4
When “END” is displayed, press the [ENT] key.
The screen will return to the axis selection screen.
If you wish to continue with checking the X axis, go
back to step 3 for calibration check.
5
When you have completed the adjustment, turn off
the power.
After adjustment is completed, go through the
checking procedure to check if the adjustment was
done accurately.
If the correction value calculated exceeds the allowable range, the RL-200 2S will display
error code [CALIBRATION OVER ERR]. Check the procedure again and perform any
inspections and adjustments.
If this error code is displayed again, repair is required. Contact your dealer or Topcon.
After positioning the laser beam, press the [ENT]
key.
After calibration value is fixed, press the [ENT]
key.
The screen goes back to the axis selection
screen.
CALCULATING
END
X Y
W X