20
SPECIFICATIONS AND PERFORMANCE
SPECIFICATIONS AND PERFORMANCE
PERFORMANCE
• The cornea fluorescent filter is not available.
• The specification and design of the product can be altered for improvements without prior notice.
Agency Compliance
The SL-2G is designed to comply with the following agency standards:
• IEC 60601-1: 1988+A1:1991+A2:1995
• EN60601-1: 1990 +A1:1993 +A2:1995 +A13:1996
• IEC 60601-1-2 Ed3.0: 2007/EN60601-1-2: 2007
• UL 60601-1: 2003
• ISO 10939:2007
• ISO 15004-1:2006
• ISO 15004-2:2007
• EC Medical Device Directive 93/42/EEC
Microscope unit
Type
Galileo type convergent binocular tubes
Magnification
Drum, 3-step magnification
Magnification steps
10, 16, 25
Overall magnification
(actual vision field)
10.00 (
φ
22.5mm)/15.98 (
φ
14.1mm)/25.53 (
φ
8.8mm)
Eyepiece lens
Magnification: 12.5x
Binocular tubes
Diopter adjustment range: -5D to +3D
PD adjustment: 55 - 78mm
Illumination unit
Illumination field
Slit width:
0-14mm, can be adjusted gradually (14mm=circle)
Slit length:
1-14mm, can be adjusted gradually (14mm=circle)
Aperture diameter:
φ
0.3, 1, 5, 10, 14mm 1-14mm, can be adjusted gradually
Slit direction
Vertical to horizontal can be adjusted gradually
Filter
Blue filter, red-free filter
Light source
LED
Base unit
Forward-backward movement 90mm
Right-left movement
100mm
Vertical movement
30mm
Fine movement forward-back-
ward/right-left
12mm
Chinrest unit
Vertical movement
80mm
The following statement is the "Essential performance" provided for by IEC60601-1.
1) The LED light source should not go off.
2) There are no component failures.