25
SPECIFICATIONS AND PERFORMANCE
Recommended separation distance between portable and mobile RF com-
munications equipment and the SL-2G
The SL-2G is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the SL-2G can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communica-
tions equipment (transmitters) and the SL-2G as recommended below, according to the maxi-
mum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance
d
in
meters
(m) can be estimated using the equation applicable to the frequency of
the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
d
1.2 P
=
d
1.2 P
=
d
= 2.3 P