- 92 -
4.1.38 CXL Command
If this command is sent during measurement, the instrument stops the immediate measurement and
sends an error code.
Measurement data during measurement is not sent. If the measurement is not in progress, a
reception confirmation command is sent.
①
Measuring
PC
Command communication direction
Measuring instrument
“ST”+(Cr•Lf)
“OK”+(Cr•Lf)
Measurement starts.
• • • • • •
“CXL”+(Cr•Lf)
Measurement stops.
“E002”+(Cr•Lf)
“END”+(Cr•Lf)
If this instrument receives this command while measuring data is being sent, it will send a
confirmation command after measurement data has been sent. Commands are not accepted
during measurement data transmission.
②
Transmitting measurement data
PC
Command communication direction
Measuring instrument
“ST”+(Cr•Lf)
“OK”+(Cr•Lf)
Measurement starts.
• • • • • •
Measurement is ended.
Data communication
Measurement data 1+(Cr
•Lf)
Measurement data 2+(Cr
•Lf)
“CXL”+(Cr•Lf)
Measurement data 414+(Cr
•Lf)
“END”+(Cr•Lf)
“OK”+(Cr•Lf)
Note