- 80 -
4.1.11 DR[n] Command
The local measurement is completed and the measurement data stored by the instrument is read
out by the computer.
“n” is specified with 1 ~ 50.
☞
‘1.3.2 Connecting the Personal Computer’
☞
‘4.2.2 Output Format for Reading out the Internal Stored Data’
PC
Command communication direction
Measuring instrument
“DRn” + (Cr•Lf)
“OK”+(Cr•Lf)
Data No. + (Cr
•Lf)
“380”+(Cr•Lf)
• • • • • • •
“780...” + (Cr•Lf)
“END” +(Cr•Lf)
4.1.12 KO1/KO2/KN1/KN2 Command
Change the use of the correction factors in the instrument.
KO1 : Changes the setting to use the correction factor for spectrum in the instrument.
KO2 : Changes the setting to use the correction factor for tristimulus values in the instrument.
KN1 : Changes the setting not to use the correction factors for spectrum in the instrument.
KN2 : Changes the setting not to use the correction factors for tristimulus values in the
instrument.
PC
Command communication direction
Measuring instrument
KO1/KO2/KN1/KN2+(Cr
・
Lf)
OK+(Cr
・
Lf)