- 87 -
4.1.28 FX Command
Starts the measurement to calculate the optimum inner-filter and integral time setting in FIX INTEG
mode. When the measurement complete, the optimum inner-filter type and integral time is retuned
and the values are also set on the instrument.
(
SR-LEDW only
)
PC
Command communication direction
Measuring instrument
FX + (Cr
・
Lf)
OK + (Cr
・
Lf)
Integral time+ (Cr
・
Lf)
Inner- (Cr
・
Lf)
END” + (Cr
・
Lf)
Remember!
This measurement should be conducted at the same condition as the actual
measurement in measurement angle, object and measurement distance.
Remember!
The integral time and inner-filter setting in FIX INTEG mode are erased
when the power off.
4.1.29
FO Command
Sets to FIX INTEG mode
(
SR-LEDW only
)
This setting is stored even after the power off
PC
Command communication direction
Measuring instrument
FO+(Cr
・
Lf)
“OK”+(Cr
・
Lf)
”
+Cr & Lf