- 40 -
3.
The scale will now display second warning indicating that all data will be permanently lost. To abort
using the navigation key select
“NO”,
or select
“YES”
to proceed and press ENTER.
WARNING: Proceeding will result in formatting and erasing the entire memory bank. All data will be
permanently lost.
4.
After proceeding with memory erasure the scale will begin to delete and format the entire database.
This process lasts up to 45 seconds as the memory is formatted.
8.4.
Calibration
The scale will automatically calibrate itself at start up. Calibration should be also performed periodically
or whenever the scale is moved or relocated. To initiate internal calibration at anytime follow the steps
below.
1.
Make sure the pan is free from any weight and press the calibration key (8).
2.
The scale will display a command indicating to confirm internal calibration.
3.
In order to confirm internal calibration and proceed, press the calibration key again. The scale will
begin internal calibration.
Calibration can last 30 to 50 seconds, when finished the scale will automatically return to the weighing
mode.