9.10. SQC Process Control
1.
Press the FUNCTION key to access the limit set database.
2.
Use the navigation key to select the
“Desired Limit Set
” and press the ENTER key to confirm your
selection.
3.
If required, place a container on the pan and press the “T” key to tare.
4.
Place the first sample on the pan. Once the weight stabilizes press the ENTER key to accept the weight
of the sample.
Note: The number of required samples is displayed in the upper left corner of the display.
5.
Remove the sample from the pan and weigh the next sample as required.
6.
Repeat the steps until all samples are entered.
7.
After entering the last sample an SQC Report will be displayed. To print the report press F1. To view
statistical data such as standard and relative deviation, as well as min and max values press F3. To exit
and start a new SQC Transaction press F5.