The front panel power switch is also
a circuit breaker. As a circuit breaker,
it prevents excessive current from
entering the PIU.
When the power is on, the switch is
illuminated. When the breaker trips,
the
switch
returns
to
its
“off”
position.
Front Panel Layout
Circuit Protection
Illuminated Power Switch / Circuit Breaker
Torus PIUs will shut-down if internal
unit temperature reaches excessive
levels.
Thermal Protection
Electrical Specifications
Input Voltage
Nominal
120VAC, 60Hz
Range +/- 10% voltage
57-63Hz
Model Number
RM 2.5
Input
Current
Limiting
2.5A
Circuit Breaker
Front Panel
Input Current
Rating
2.5A
½ cycle
25A
Load Current Capability
1 second
20A
10 seconds
.
7 5A
www.TorusPower.com
Owner’s Manual
Audio / Video Power Isolation Units
Rack Mount / Standalone
10/07
Power Conditioners
™
l
O
T
oronto
,
Canada
www.TorusPower.com
Model Number
Input
Output
Part Number
Serial Number
RM 2.5
120VAC 60Hz, 2.5A max
120V 2.5A max
PA-10201-00
0001
WARNING
DISCONNECT POWER
BEFORE WORKING WITHIN
Torus PIUs are high-power products.
The outlets they are plugged into
should be sufficient to provide the
current to operate them without
tripping circuit breakers within your
junction panel.
Rear Panel Layout
Wall Receptacles
Product Label
Power Inlet
Accessory
Ground Stud
Power Outlet
Mechanical Specifications
Model Number
RM 2.5
Input
Connector
(Rear Panel)
IEC 15A
Inlet
Line Cord
(included)
N5/15, 14AWG
C13, 15A/125V
Suggested
Wall
Receptacle
Standard outlet
NEMA 5-15R
15A - 125V
Output
Connector
(Rear Panel)
Qty 2 (4 outlets)
Medical-grade
Weight
17 lbs.
Size
(w” x d” x h”)
19.0 x 11.0 x 2.0
With mounting feet
19.0 x 11.0 x 1.75
Without mounting feet
Warning Label
Removable Mounting
Feet for Rack Mount