10
Power distribution board
Ambient conditions
Temperature
Humidity
Heat generation
Scan room
Gantry
20°C to 26°C
40% to 80%
Approx.
Tolerance: ±2°C No
9.720 kJ/h (*1)
condensation 32,070 kJ/h (*2)
Patient couch 20°C to 26°C
40% to 80%
Approx.
Tolerance: ±2°C No
1,080 kJ/h (*1)
condensation 1,800 kJ/h (*2)
Operator’s room
Console
16°C to 28°C
40% to 80% Approx.
No 9,720
kJ/h
condensation
*1: When scanning is not performed.
*2: When scanning is performed continuously at the maximum
rated output of the system.
Minimum area for installation
• For the long patient couch version:
22 m
2
• For the short patient couch version:
21 m
2
NFB
150 A
NFB
100 A
In case of
3-phase, 200 V
In case of
3-phase, 400 V
System
System
transformer
(option)
Ground resistance:
As per applicable
legal requirements.
Ground bar
Room layout example
Unit: mm
1
2
3
1,600
3,072
Gantry
Patient couch
*: For the short patient couch version
Console
5,530 (4,930*)
7,130 (6,530*)
(1,194*)
1,494
1
2
3