background image

4

SURE

Start

TM

:

– Continuous scan time: Max. 100 s
– Region of interest 

(ROI):

Max. 3 ROIs

– CT number 

measurement interval: 0.083 s (12 measurements/s)

– Scan start delay time: Min. 3 s
– Display function:

Mean CT number and elapsed
time

• Specification of

reconstruction position: By entering the couch-top posi-

tion or using the scanogram 

• Reconstruction method: TCOT reconstruction (applicable

to 16-row acquisition data)
MUSCOT reconstruction (applic-
able to 4-row acquisition data)

• Reconstruction mode:

Full image
Half image
Detail image

• The relationships between the slice thickness and the

imaging area scanned in 7.5 s (0.75 s, 10 rotations) in
the longitudinal direction are shown for helical pitches 
10 and 20.

*: The couch-top traveling distance per rotation is shown rela-

tive to the slice thickness.

Dynamic scan

• Scan time:

0.75, 1, 1.5 s/360°

• Programmable time:

Max. 1 hour 
This refers to the maximum time
within which a series of scans is
performed following a predeter-
mined eXam Plan.

• Number of 

programmable scans:

Max. 10
Max. time of one continuous
scan is 100 s.

• Scan plan

– Scan interval:

Min. interval is 1 s 
Setting is possible in increments
of 0.1 s in a scan interval of
more than 1 s.

Note: When a scanning mode with patient couch movement is

used, the minimum scan interval is limited by the time
required for movement.

• Scan start delay time:

Min. 0.5 s
Setting is possible in increments
of 0.1 s.

• Scan rate:

Max. 133 scans/100 s 
(0.75-s scan, 133 rotations)

e

c

il

s

g

n

i

t

t

e

S

s

s

e

n

k

c

i

h

t

*

h

c

t

i

p

l

a

c

il

e

H

0

1

0

2

m

m

5

.

0

m

m

0

5

m

m

0

0

1

m

m

1

m

m

0

0

1

m

m

0

0

2

• Image reconstruction

– Number of images:

Max. 4 images/scan

– Image interval:

Reconstruction is possible in
increments of 0.1 s.

• Reconstruction time:

Min. 0.5 s

• Real-time reconstruction 

time:

12 images/s (0.083 s/image)
(1 slice, 512 

×

512 matrix)

X-ray generation

• X-ray exposure:

Continuous

• X-ray tube voltage:

80, 100, 120, and 135 kV

• X-ray tube current:

10 mA to 300 mA (260 mA for

135 kV) (in increments of 10 mA)

• X-ray tube heat capacity:4.0 MHU
• X-ray tube cooling rate: Max. 864 kHU/min 
• Focal spot size

– IEC 60336: 1993

nominal :

0.9 mm 

×

0.7 mm (small)

1.4 mm 

×

1.4 mm (large)

X-ray detection

• Detection system:

Solid-state detectors

• Main detector:

800 channels 

×

28 elements

• Data acquisition:

800 channels 

×

16 rows

• Reference detector:

1 set

• View rate:

Max. 1,200 views/s

Data processing

• Reconstruction matrix:

512 

×

512

• Picture element (pixel) size

– CT image:

Unit: mm

Scan field

SS

S

M

L

LL

Pixel size

* to

* to

* to

* to

* to

0.35

0.47

0.63

0.76

0.98

*:  Depending on the Vari-Area or Zoom factor

– Scanogram:

Unit: mm

• Reconstruction filter functions

– Functions for the abdomen with BHC
– Functions for the abdomen without BHC
– Functions for the brain with BHC
– Functions for the brain without BHC
– Functions for the inner ear and bone
– Functions for the lung
– Functions for high-resolution mode
– Functions for super-resolution mode for the inner ear,

bone, and lung

– Functions for maintenance

Standard

Enlargement ratio (area)

4.00
2.00

1.0

0.50

LL

L

M

S

Pixel size

Summary of Contents for Activion 16

Page 1: ... data acquired using thin slices images with reduced partial volume effects can be obtained Outstanding operability Operability is improved as described below In 3D image processing or time consuming image pro cessing such as for regions in which calcified areas are superimposed on contrast medium bone elimination can be performed easily while observing reference images Most 3D images can be gener...

Page 2: ...r synchronization system CKIS 004A Orbital synchronized scan system CKOS 001A Pediatric scanning system CHKS 002A Not available in the U S A PERFORMANCE SPECIFICATIONS Scan parameters Scan regions Whole body including head Scan system 360 continuous rotate rotate Scan plan programming More than 360 different sequences can be pre pro grammed Scan time CT scan 0 48 s Partial 0 75 1 1 5 2 and 3 s Sca...

Page 3: ...image Real time helical reconstruction time 12 images s 0 083 s image 1 slice 512 512 matrix MPDCT0290EA Patient couch Vertical movement System Hydraulically driven Speed of vertical movement Up 16 to 24 mm s 50 Hz 19 to 28 mm s 60 Hz Down 20 to 30 mm s Stroke Approx 588 mm Minimum couch top height Approx 310 mm Maximum couch top height Approx 944 mm Couch top movement System Motor driven or manua...

Page 4: ...can rate Max 133 scans 100 s 0 75 s scan 133 rotations e c i l s g n i t t e S s s e n k c i h t h c t i p l a c i l e H 0 1 0 2 m m 5 0 m m 0 5 m m 0 0 1 m m 1 m m 0 0 1 m m 0 0 2 Image reconstruction Number of images Max 4 images scan Image interval Reconstruction is possible in increments of 0 1 s Reconstruction time Min 0 5 s Real time reconstruction time 12 images s 0 083 s image 1 slice 512 ...

Page 5: ...unction Software control function key Multi frame display Reduction cut off display ROI processing Inset scanogram display Selective related information display Cine display Image display speed Variable Scanogram CT image switching Show scano line zoom hide scano line Slice feed playback CineView High speed image feeding using the mouse or keyboard Image processing Scanogram processing Slice posit...

Page 6: ...U TIFF conversion Filming Ethernet Toshiba protocol DICOM PRINT SCU Sheet editing function using virtual film T mode Related information items such as the patient name are dis played in the footer area using a larger font Note To use T mode the laser imager must support 2048 pixels x 2404 pixels for a 1 x 1 frame Auto filming in eXam Plans IMAGE QUALITY Noise Standard deviation Less than 0 5 Scan ...

Page 7: ...built into the gantry and the system employs a high frequency inverter for generat ing and stabilizing the high voltage supplied to the X ray tube The generator includes electronic circuits for control ling the speed of the rotating anode in the X ray tube Use of a high frequency inverter system results in high power output combined with excellent stability In addition the system is compact and li...

Page 8: ...for rapid patient scanning during a single breath hold and for high defini tion three dimensional and MPR imaging Real time helical reconstruction mode makes it possible to observe the images being scanned in real time at a maximum at 12 frames per second This mode shows any shift in the slice position in real time and helps the operator to check the scan field on the image the con trast study tim...

Page 9: ...er scan Real time reconstruction is possible in scanoscopy CT images can be reconstructed in 0 5 second for 0 75 second routine scans Ease of operation is ensured by incorporating a hybrid keyboard mouse driven menus and a large color LCD screen The couch top can be lowered very near the floor simpli fying patient transfer COMPLIANCE Council Directive 93 42 EEC Concerning Medical Devices Medical D...

Page 10: ...nsation 1 When scanning is not performed 2 When scanning is performed continuously at the maximum rated output of the system Minimum area for installation For the long patient couch version 22 m2 For the short patient couch version 21 m2 NFB 150 A NFB 100 A In case of 3 phase 200 V In case of 3 phase 400 V System System transformer option Ground resistance As per applicable legal requirements Grou...

Page 11: ...quired for routing cables that connect the various units The operator s room should have entrances for access to the corridor and the scan room Cable connections between units Checks before bringing in the unit Check in advance the width of the corridor the dimen sions of the entrance and the dimensions and maximum allowable load of the stairs and elevators to ensure that it is possible to bring i...

Page 12: ...12 OUTLINE DRAWINGS Gantry and Patient Couch ...

Page 13: ...MPDCT0290EA 13 OUTLINE DRAWINGS 344 13 5 556 21 9 1 910 75 2 900 35 4 2 070 81 5 TILT ANGLE Gantry ...

Page 14: ...OUTLINE DRAWINGS 1 862 73 3 2 690 105 9 588 23 1 1 038 40 9 STROKE 450 17 7 2 187 86 1 2 190 86 2 470 18 5 630 24 8 430 When the arm up holder is mounted 17 Patient Couch for the long patient couch version ...

Page 15: ...EA 15 OUTLINE DRAWINGS 1 562 61 5 2 390 94 1 588 23 1 1 038 40 9 STROKE 450 17 7 1 887 74 3 1 890 74 4 470 18 5 630 24 8 17 430 When the arm up holder is mounted Patient Couch for the short patient couch version ...

Page 16: ...of Toshiba Medical Systems Corporation This document may include trademarks and registered trademarks of other companies Model TSX 031A 2007 04 TME KI Toshiba Medical Systems Corporation Nasu Works meets the Environmental Management System standard ISO 14001 Toshiba Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001 ISO 13485 Printed in Ja...