No. 2A201-519EN*A
11
(8) External dimensions and mass
External dimensions
Mass
Unit name
Width (mm)
Depth (mm)
Height (mm)
(kg)
Gantry (with cover)
2,330
960
1,950
1,750
Gantry (without cover but with
median-line metal plate)
2,290 875 1,935 1,650
Patient couch (CBTB-019A)
630
2,690
450
480
Patient couch (CBTB-019B)
630
2,390
450
450
Console (CKCN-012B/2A)
*1
2,470
815
1,470
435
Computer system (CPU BOX)
450
815
700
140
(REC
BOX)
870
650
1,470
260
Speaker 140
135
200
2
System transformer
800
770
980
550
*1
: The dimensions of the console are those when a recommended table is used.
(9) Maximum cable length
See subsection 8.6 "Cable Connections Between the Units".
(10) Installation
altitude
The installation altitude should be no more than 1000 m above sea level. Prior consultation is
required for installations higher than 1000 m.
(11) Transport and storage conditions
Observe the following conditions during transport and storage after shipment.
Ambient temperature : -10
°
C to 50
°
C
Relative humidity
: 40% to 80% (No condensation)
Atmospheric pressure : 700 hPa to 1060 hPa
Vibration :
9.8
m/s
2
(1G) or less (during storage)
: 19.6
m/s
2
(2G) or less (during transport)