No. 2A201-519EN*A
14
(4) Miscellaneous
The following should be provided where required:
•
Washstand
: For water phantom or contrast medium injector
•
Storage rack
: For first-aid kit or accessories
•
Gases required for medical treatment : Oxygen and nitrogen dioxide
•
Vacuum
•
Ceiling traveling irrigator
: For instillator
(5) Air
conditioning
Since the gantry includes components such as detectors or electronic circuits which permit
high-precision measurement, air conditioning is required to maintain high precision. For
details, refer to subsection 9.1 "Air Conditioning".
(6) Illumination
Illumination should be adjustable, because it is easier to specify the desired scan slice
position using the positioning projectors when illumination is decreased.
(7) Wiring pits and ducts
The wiring pits and ducts should be installed in accordance with subsection 7.5 "Wiring Pits
and Wall Ducts".