No. 2A201-519EN*A
21
7.
BUILDING CONSTRUCTION WORK
7.1
Entrance Used to Bring-in the Unit
To carry the units into the room without any difficulty, conditions such as sufficient door size,
corridor width and ceiling height must be satisfied in advance.
(1) Bringing-in
method
The system is normally shipped with the covers mounted. If it is necessary to disassemble
the system to reduce the external dimensions or mass for any reason, consult Toshiba in
advance.
(2) External dimensions and mass
The following figure shows the external dimensions and mass of the gantry and the required
size of the bringing-in entrance.
Recommended values
Height of the door
2050 mm or more
Width of the door
1100 mm or more
Load of the elevator
2 tons
Mass: 1720 kg
Figure 7.1-1
(3) Carrying-in
jig
To install the main body of the scanner easily, it is recommended that the carrying-in jig
should be used.
Mass: 1750 kg