No. 2A201-519EN*A
33
8. ELECTRICAL
WORK
8.1 Electrical
Equipment
8.1.1 Electrical
specification
Application
For X-ray system
(gantry rotation section)
Gantry stationary
section/console
Number of phases
Three phases
Line voltage
200 V
Frequency
50 Hz or 60 Hz
Power capacity
*1
100
kVA
Rated capacity
(Actual load capacity)
85 kVA
14 kVA
Voltage variation due to
changing load
5% or less
Line voltage variation
*2
±
10% or less
(*1) The cables between the power receiving transformer and the Aquilion power distribution
board must exceed the minimum nominal sectional area, obtained from the cable length
and the transformer capacity, as shown in figures 8.1-1 and 8.1-2.
In addition, independent cabling is desirable (the cable for the X-ray system (gantry
rotation section) should be separate from the cable for the gantry stationary
section/console).
(*2) This specification must be met even when a load is applied.