No. 2A201-519EN*A
44
Table 8-1 List of inter-unit cables (1/3)
Power cables
Cable
No.
Wiring
pit
Connected
to: (1)
Connected
to: (2)
Standard
length (m)
Effective
length (m)
Retract length
(m)
Maximum
length (m)
Number of cores
×
section area (mm
2
)
Outer cable
diameter (mm)
Connector diameter (mm)
Remarks
P01 P System
trans.
Console 20
(10)
18
(8)
1.5 while
connecting to (1)
0.5 while
connecting to (2)
20
3 cores
×
5.5
17.5 Solderless
contact
while
connecting to (1)
Solderless contact while
connecting to (2)
P02 P System
trans.
Gantry 15
(11)
12
(8)
1.5 while
connecting to (1)
1.5 while
connecting to (2)
15
3 cores
×
5.5
17.5 Solderless
contact
while
connecting to (1)
Open wire while connecting
to (2)
P04 P Gantry Console
CPU BOX
19
(14)
17
(12)
1.5 while
connecting to (1)
0.5 while
connecting to (2)
19
6 cores
×
0.75
11.5 Solderless
contact
while
connecting to (1)
Connector (28
×
21
×
28)
plus solderless contact
P00 P Power
distribution
board
System
trans.
5 3
1.5
while
connecting to (1)
0.5 while
connecting to (2)
7
3 cores
×
35
35.2 Solderless
contact
while
connecting to (2)
E01 P Power
box
Gantry
15
(11)
12
(8)
1.5 while
connecting to (1)
1.5 while
connecting to (2)
15
1 core
×
35
12.2 Solderless
contact
while
connecting to (1)
Solderless contact while
connecting to (2)
P03 P System
trans.
Gantry 15
(11)
12
(8)
1.5 while
connecting to (1)
1.5 while
connecting to (2)
15
3 cores
×
35
35.2 Solderless
contact
while
connecting to (1)
Exposed cable at (2)
E00 P Power
box
System
trans.
5 3
1.5
while
connecting to (1)
0.5 while
connecting to (2)
7
1 core
×
35
12.2 Solderless
contact
while
connecting to (2)