No. 2A201-519EN*A
45
Table 8-1 List of inter-unit cables (2/3)
Signal cables (1/2)
Cable
No.
Wiring
pit
Connected
to: (1)
Connected
to: (2)
Standard
length (m)
Effective
length (m)
Retract length
(m)
Maximum
length (m)
Number of cores
×
section area (mm
2
)
Outer cable
diameter (mm)
Connector diameter
(mm)
Remarks
S01 S Console
REC BOX
Gantry 19
16
0.5
while
connecting to (1)
1.5 while
connecting to (2)
19 2-pairs,
optical
fiber
11
Connector (22
×
8
×
25) Minimum
permissible
bending
radius:
110 mm
S02 S Console
CPU BOX
Gantry 19
(14)
16
(11)
0.5 while
connecting to (1)
1.5 while
connecting to (2)
19
18 pairs
×
0.08
shielding
7.1
Connector (62
×
20
×
52) while connecting to
(1)
Connector (40
×
12
×
42) while connecting to
(2)
(a)
Gantry 19
(14)
16
(11)
0.5 while
connecting to (1)
19
3 pairs
×
0.18
shielding
4.2
Connector (
φ
29
×
52)
while connecting to (1)
S03 S Console
CPU BOX
(b) Speaker
(Gantry)
19
(14)
16
(11)
2.0 while
connecting to (2)-
(a)
19
2 pairs
×
0.5
shielding
5.1
Connector (
φ
25
×
49)
while connecting to (2)-
(a)
(c)
Speaker
(C-room)
11 8
2.0
while
connecting to (2)-
(b)
2.0 while
connecting to (2)-
(c)
11
2 pairs
×
0.5
shielding
5.1
S04 S Console
REC BOX
Gantry 14
11
0.5
while
connecting to (1)
1.5 while
connecting to (2)
19 2-pairs,
optical
fiber
11
Connector (22
×
8
×
25) Minimum
permissible
bending
radius:
110 mm