No. 2A201-519EN*A
69
Figure A.1.2-3 illustrates an example of walking in place followed by standing on a
conductive floor with normal shoes and conductive shoes. Preventive measures
against static electricity using a conductive floor and conductive shoes are extremely
effective.
This method should therefore be used during site planning to prevent damage due to
static electricity caused by users or service personnel.
Materials and manufacturers are listed below.
Reference manufacturer: Achilles Inc.
•
Conductive floor ------------------ Conductive floor tiles SKY-11
•
Conductive slippers ------------- ICS-0100
•
Conductive shoes --------------- ICM0100 (for men)
ICM0230 (for women)
The installation area of the conductive floor is shown in the attachment. (Refer to
figure A.1.2-4 "Room layout example".)
It is recommended that special consultation with the flooring material manufacturer be
held before building the floor.