No. 2A201-519EN*A
70
Conductive floor space
(1) GTS side service space
Connect the ground cable of the
conductive floor to the gantry.
(2) Servo amplifier side service
space
Connect the ground cable of the
conductive floor to the gantry.
(3), (4)
Gantry front service space
Connect the ground cable of the
conductive floor to the patient
couch.
(5) Gantry rear side service space
Connect the ground cable of the
conductive floor to the gantry.
(6) Console (REC BOX) service
space
Connect the ground cable of the
conductive floor to the console
(REC BOX) ground.
(7) Console (control room) service
space
Connect the ground cable of the
conductive floor to the console
ground.
Unit: mm
1600
800
26
90
2100
(7)
(3)
(5)
(1)
(2)
(4)
(6)
Figure A.1.2-4 Room layout example