e-STUDIO350/352/353/450/452/453
© 2003 - 2008 TOSHIBA TEC CORPORATION All rights reserved
ADJUSTMENT
3 - 102
(5) Press the [Button1] to adjust the alignment position.
Every time the [Button1] is pressed, the alignment plate shifts 1 step (0.419 mm/step) toward the
“+ “direction. (The gap between the alignment plates becomes narrower.)
Adjustment range is from -5 to +5 steps.
If the [Button1] is pressed when the alignment position is at the “+5 step”, the plate will return to
the home position and then moves to the position of “-5 step”.
Fig. 3-140
(6) When the adjustment is completed, press the [Button2] on the finisher control panel to store the
adjustment value in memory.
When the value is stored normally, the [LED1] on the control panel will blink for a number of times
that corresponds to the adjustment value set for the equipment.
See the following table for the number of times the [LED1] blinks and its corresponding adjust-
ment value.
(7) Turn OFF the power of the equipment.
(8) Turn OFF all bits of the SW1 on the Finisher control board.
(9) Install the board access cover.
Number of Blinking
Adjustment Value
1
-5
2
-4
3
-3
4
-2
5
-1
6
0
7
+1
8
+2
9
+3
10
+4
11
+5
LED3
LED2
LED1
Button2
Button1
06/09