6F8A0917
- 132 -
10.3
Multi-Range
Functions
Five types of multi-range functions can be set using the Range Type:
Range
Typ
(1) Single direction, 4 ranges
Internal automatic switching
(2) Forward and reverse directions, 2 ranges
Internal automatic switching
(3) Single direction, 2 ranges
External signal switching
(4) Forward and reverse directions, 2 ranges
External signal switching
(5) Single direction, 4 ranges
E xternal signal switching
To use the multi-range function, proceed as follows to set.
Range Setting
•
Follow the procedure in 8.2.8 “Span Value (Range)” and set the range in the following
order.
1. Select the Range Type.
2. Set the span for Ranges 1 to 4.
3. Set the hysteresis value.
DO/DI Setting
•
Follow the procedure in 8.2.14 “Digital I/O Function” and set the DO for Range output.
•
When using the multi-range switching with external signals, set the DO for external
range switching signals.
•
Requirements to combine the range signals (No.1 and No.2) to set the DO and DI differ
depending on the Range Type. For details, see the explanations (1) to (5) on the
following pages.