6F8A0917
- 166 -
3. Other specifications
•
Conductivity:
5
μ
S/cm or more
(If detector is partially filled pipes type, conductivity is 100
μ
S/cm or more)
•
Power supply:
100 to 240VAC (allowable voltage range: 80 to 264VAC, 50/60Hz)
*1
or
24VDC (allowable voltage range: 21 to 27VDC)
*2
•
Power consumption: 17W (27VA ) or less (when 7th digit of specification code is “A”)
35W (60VA ) or less (when 7th digit of specification code is “B”)
30W (50VA ) or less (when 7th digit of specification code is “F”)
*1
When the 7th digit of specification code is “B” and “F”, power supply is 100 to 120VAC
(allowable voltage range is 80 to 132VAC, 50/60Hz)
*2
When the 7th digit of specification code is “B” and “F”, 24VDC cannot be selected.
•
For partially filled pipes detectors:
Fluid water level range:
Meter size 6”-12” (150-300mm) : Water level 1 1/4(30mm) to state to be filled with water
Meter size 14”-24” (350-600mm) : 10% of meter size to state to be filled with water
Note) State to be filled with water is water level 100%
Necessary straight pipe length:
Upstream side : 10D or more
Downstream side: 5D or more
Note) D is meter size.
Converter specifications
1. Input signal: Flow rate proportional signal from the detector
2. Output signal: 4 to 20mADC (current signal, load resistance 0 to 1k
Ω
)
3 Digital Input/Output
•
Digital Output
Digital Output 1: Output type
Transistor open collector
Number of outputs 1
Capacity 30VDC,
200mA
max.
Digital Output 2 (option: added when 9th digit of specification code is “3”) :
Output type
Solid-state contact output
Number of outputs 3
Capacity 150VDC,
150mA
max.
150VAC (peak value), 100mA max.
* For digital output function, either of the following functions can be selected.
Digital Output function
•
Totalizer pulse output
DO1 or DO2 can be selected
(Forward direction pulse and reverse direction pulse can be assigned independently)
In the case of DO1 Pulse rate 3.6 to 3,600,000 pulses/h
Pulse width
Settable within the range of 0.3 to 500ms
In the case of DO2 Pulse rate 3.6 to 360,000 pulses/h
Pulse width
Settable within the range of 4 to 500ms