6F8A0917
- 20 -
(4) If the converter needs to be installed in a location where watertight installation is required, make
unused cable ports watertight. (Be careful that the attached blind plate is used for dustproof
purpose and it is not effective for watertight installation.)
(5) The converter has a surge arrestor/protector installed inside. Therefore, do not conduct a withstand
voltage test for the converter. In addition, to check the insulation of the converter, use a voltage of
250VDC or less.
(6) When wiring is completed, be sure to install the protection cover of the terminal block.
5.1 Installation Cables
Use the kind of cables shown in Table 5.1 to connect to the converter.
Table 5.1 Installation Cables
Name Cable
name
Nominal
cross-sectional area
Overall diameter
Description
Flow rate signal
cable
2-core shielded
chloroprene cabtyre
cable
( Rubber covered cable )
0.75mm
2
0.433
−
0.512 inch
(11
−
13mm)
JIS C 3327 or
equivalent
Excitation cable 3-core chloroprene
cabtyre cable
( Rubber covered cable )
2mm
2
1.25mm
2
0.433
−
0.512 inch
(11
−
13mm)
JIS C 3327 or
equivalent
Power cable
3-core vinyl sheathed
cable or 2-core vinyl
sheathed cable
2mm
2
0.433
−
0.512 inch
(11
−
13mm)
CVV JIS C 3401 or
equivalent
Output signal
cable
The number of insulated conductors the cable contains differs
depending on the specification of the output signal cable.
Use a shielded cable of overall diameter 0.433 to 0.512 inch
(11 to 13mm) with nominal cross-sectional area of 1.25mm
2
.
CVV-S
JCS-258-C or
equivalent