5.4 Wiring Method
Do not conduct wiring work when
power is
applied
.
Do not work on piping and wiring
with wet
hands
.
DON’T
Wiring while power is applied can
cause
electric shock
.
DON’T
Wet hands may result in
electric
shock
.
5.4.1 Terminal Treatment of Cables
Proceed as follows to treat the terminals at the converter side of the flow rate signal and excitation
cable and to connect these cables to the terminal block. Use appropriate cables referring to 5.1
"Installation Cables." Attach and crimp a round type M4 insulated crimping terminal to the end of each
cable.
Excitation cable
Strip the sheath from the end of each wire as shown in Figure 5.2 and attach and crimp an M4
crimping terminal with insulated sleeve to the end of each wire and then connect the crimped
terminals to X and Y of the terminal block. Connect the terminal of red wire to E of the terminal
block.
X Black
E Red
Y White
M4 crimped terminal
25 to 45mm
Figure 5.2 Terminal Treatment of Excitation Cable