Strata CT Installation
2-52
Strata CT I&M Manual October 2000
8.
Connect the fiber optic cables
◆
Pass the fiber optic cable through the protective tube.
◆
Route the tube through the clamp attached to the inner cabinet wall and secure the
clamp
◆
Attach fiber to the SC connectors on the ROMS1A daughterboard.
◆
The TX side of the Master connects to the RX side of the slave.
◆
The RX side of the Master connects to the TX side of the slave.
◆
TAIS recommends that the cables be marked within the cabinet for ease of
maintenance.
◆
Observe the minimum bend radius of 30mm.
9.
Restore power.
10. Run Command Program 91-1 Automatic PCB Recognition and Port Renumber.
The RRCU PCB and its controls and connectors are shown in
.
Figure 2-34
Remote Expansion Cabinet Printed Circuit Board (RRCU1A)
Table 2-15
Remote Cabinet Data Cables and Connectors
Data Cables
RRCU Connectors
M1
S1
M2
S2
BDCL1A-MS1
X
X
BDCL1A-M2
X
BDCL1A-S2
X
X
=applies to connector
4356
RRCU1A
Backplane Connector
P8
P9
P7
P6
M2
S2
S1
M1
Master 1
Connect to
Base Cabinet
Master 2
Connect to
Base Cabinet
Slave 1
Connect
to MSBU
Slave 2
Connect
to MSBU
RS-232
Port
ROMS1A
MS
MS
P11
P10
P4
P3
P2
P1
H
C
Summary of Contents for Strata CT
Page 36: ...Strata CT Configuration 1 22 Strata CT I M Manual October 2000 H C ...
Page 94: ...Strata CT Installation 2 58 Strata CT I M Manual October 2000 H C ...
Page 242: ...Peripheral Installation 6 54 Strata CT I M Manual October 2000 H C ...
Page 258: ...ACD Installation 7 16 Strata CT I M Manual October 2000 H C ...
Page 274: ...Fault Finding 8 16 Strata CT I M Manual October 2000 H C ...
Page 364: ...ISDN Interfaces 10 38 Strata CT I M Manual October 2000 H C ...
Page 370: ...Notes to Users A 6 Strata CT I M Manual October 2000 H C ...