Toshiba VP Systems Release 7.1
July, 1994
MAINTENANCE
5-57
Table 5-3 Hardware Error Types and Remedies (continued)
Error Type:Name
Source
Remedy
Priority
11: Disk Initialization
Failure
VCU could not
successfully initialize
the disk(s)
Contact Toshiba Technical Support.
3
Parity error occurred
in CPU memory
If error repeats, replace power
supply unit.
3
12: CPU PARITY
DATA1 =
Reason:
01
02
NOT USED
NOT USED
NOT USED
=
=
NO SYSTEM DISK FOUND
NO VALID FILE SYSTEM FOUND
DATA-1
DATA-2
DATA-3
DATA-4
FROM I/O ADDRESS 6A (ERR.ADDR2)
BIT 0, 1, 2 = CONTEXT REG IN USE ON ERROR
FROM I/O ADDRESS 69 (ERR.ADDR1)
HIGH ADDRESS ON PARITY ERROR
FROM I/O ADDRESS 68 (ERR.ADDR0)
LOW ADDRESS ON PARITY ERROR
CONTENTS OF CONTEXT REG DESIGNATED BY DATA-1
Temperature or
voltage out of range.
Check DATA-1 for information
pertaining to what is out of range
and perform corrective action.*
15: System Status
Error
3
* To correct cabinet temperature, check the filter, fan, room
temperature, etc. Be sure there are at least six inches
between the VP 300 cabinet and the wall. If Type
15 errors (with DATA=1) still appear, replace the CPU
card. For Type 15 errors (with DATA1=2, 3, 4, 5, 6, øA, øB),
voltage cannot be adjusted, recommended action is to replace
the power supply unit.
VCU code checksum
incorrect
Contact Toshiba Technical Support
3
14: Checksum Error
DATA-2 =
DATA-3 =
DATA-4 =
center.
Summary of Contents for VP 300
Page 2: ......
Page 10: ......
Page 14: ......
Page 20: ...Toshiba VP Systems Release 7 1 July 1994 ...
Page 52: ...Toshiba VP Systems Release 6 1 February 1993 2 14 Installation and Maintenance Manual ...
Page 56: ...Toshiba VP Systems Release 7 1 July 1994 ...
Page 146: ......
Page 312: ...Toshiba VP Systems Release 7 1 July 1994 ...
Page 314: ......
Page 444: ......
Page 496: ......
Page 542: ......
Page 598: ...Toshiba VP Systems Release 6 1 February 1993 ...
Page 600: ...Toshiba VP Systems Release 6 1 February 1993 ...