XSDC 601-01
5. Installation
Personal
safety devices must be
used during this operation.
5.1 Transport
The refrigerator cabinet is shipped on a pallet and
protected by its packaging.
The refrigerator cabinet should always and only be
transported in a vertical position
(indicated Up-Down on the packaging) (see figure).
Installation and start-up must be
performed
by
qualified
and
authorized personnel only.
5.2 Environmental specifications.
When positioning the equipment, keep in mind that its operation is guaranteed only in environmental
conditions with a temperature of +18°C to +43°C.
5.3 Positioning
Check the suitability of the electrical plug installed (SHUKO) according to current national
laws. Replace the plug if not in compliance; the replacement must be done by qualifed
personnel. Check that the mains voltage corresponds with that of the REFRIGERATOR
CABINET as reported on the label. Before positioning the equipment, put on safety clothing
and gloves for moving it since the metal parts could be sharp. Position the product in a
clean, dry and well-ventilated area. The refrigerator is set for operation in environments
where the temperature is b18°C and +43°C (Tropicalized Compressor) -Climate
Class T -. The equipment should be located far from heat sources and in places with
sufficient ventilation. Do not install the equipment in places where there are explosive
and/or corrosive gases/powders. Make sure the equipment is level and on flooring that is
capable of supporting its weight and in a suitable area based on its dimensions and use.
Make sure the product is not near any heat sources.
Once the refrigerator cabinet is positioned in the intended area,
make sure it is level using the adjustable feet.
Do not locate the REFRIGERATOR CABINET near
heat sources or in areas with elevated temperatures
since this could lower performance and increase
equipment wear. Leave at least 50cm between the
equipment and any upper surface (FIG. 3)
.
Make sure the power cable is not under the
equipment.