FGH300
8
FUNCTION SETTINGS
1
.
Press [ON/OFF] to turn on the scale, the display will show flash digits and then --------,
press
[PRINT] ( this function is selectable)
, the display will show b XXXX value ( press
[UNIT]
to select among
1200,2400,4800,9600), our default setting is 9600. Press
[ZERO]
to enter Parity mode, the display will show P
XXX, press
[UNIT]
to select odd, even, none (odd stands for 7 data bits with odd parity, even stands for 7 data
bits with even parity, none stands for 8 data bit without parity)our default setting is none.
Press
[ZERO]
to enter Print mode, the display will show n XXX, press
[UNIT]
to select con, off, key, stb (OFF
stands for serial data output disabled, KEY stands for demand mode by press
[PRINT],
STB stands for
automatic print when scale is stable, con stands for continuous print, our default setting is KEY. press
[ZERO]
again to return to weighing mode.
2
.
Switch the power to “ON”
and than press
[ON/OFF]
to turn on the scale
。
The display shows F-E-9 to 0
and than “- - - - - - -”
,
press
[TARE]
key to enter ZERO track function setting
,
display shows
ZRD x.x
,
press
[UNIT]
key to select ZERO track value
(
0.0d-0.2d-0.4d-0.6d-0.8d-1.0d-1.5d-2.0d
)
our default setting is
0.8; press
[ZERO]
key to enter ZERO track time function setting
,
display shows
ZRT x.x
,
press
[UNIT]
key
to select ZERO track time
(
05. to 4.0 seconds
)
our default setting is 2.0; press
[ZERO]
key to enter ZERO
setting
,
display shows
RZD XX
, press
[UNIT]
key to select ZERO value
(
0.8d-1.8d-2.8d-3.8d
)
our default
setting is 0.8;press
[ZERO]
key to enter
FIL
setting, press
[UNIT]
key to select filter range amend (0-1-2-3-4-
5-6-7),”2”is default setting; press [ZERO] key to select zero range when turn on the scale, press
[UNIT]
key to
select percent (2,3,4,5,8,10,20) our default setting is 20; press
[ZERO]
key to set zero range, press
[UNIT]
key
to select percent (
KZR
3,4,5,10) our default setting is 4; press
[ZERO]
key to set
TARE range
, press
[UNIT]
key select percent (
KTR
50 or
KTR
100) our default setting is 50; press
[ZERO]
key to enter overload
protection setting, press
[UNIT]
key to select overload protection value (OVR 9d
OVR 2 OVR 5 OVR 10) our
default setting is 9d; press
[ZERO
] key to set beep function, press
[UNIT]
key to choose turn on or turn off the
beep, “
bz ON
” means activate the beep, “
bz OFF
” means inactivate the beep; press
[ZERO]
key to unite the
function of TARE and ZERO, press
[UNIT]
to switch between them. “
Zt ON
” means unite function activate,
“
Zt OFF
” means unite function inactivate our default setting is Zt OFF.
Press
[ZERO]
key to enter auto off function setting
,
display shows
A
xx
,
press
[UNIT]
key to select auto –
off time
(
3,5,10,off
)
our default setting is 10.press
[ZERO]
key again to enter into LED brightness function,
display shows
L x
,
press
[UNIT]
key to select brightness
(
0,1,2,3,
)
our default setting is 2.
3.