The CurveX 4 is an intelligent temperature data 
logging system that was specially designed to 
control curing processes in the coating industry. This 
manual describes how to use the CurveX 4 logger 
and how to download the measurements to your 
computer with the Ideal Finish Analysis software.

 

For the software installation instructions see the 
software manual.

 

 

4.1

 

Quick start

 

The information in the following steps should be 
sufficient to allow you to operate the CurveX 4 
without further use of this manual. For more detailed 
information we refer you to the remainder of the 
manual. See Figure 1 for more information about the 
interface positions.

 

 

Step 1

 

Set the paint type and other parameters of the 
CurveX 4 with the Ideal

 

Finish Analysis software:

 

 

Start Ideal Finish Analysis on the computer and 
connect the logger

 

 

Choose User settings in the Logger menu and 
follow the wizard. See the Ideal Finish Analysis 
help for details on the options.

 

 

Disconnect the logger from the computer

 

 

Step 2

 

Connect the thermocouple probes to the input 
points at position 7 of the

 

CurveX 4.

 

 

Step 3.

 

Attach the sensors to the object to be measured.

 

 

Step 4.

 

Press and hold down the Power button (position 1) 
for one second to switch on the

 

CurveX 4.

 

 

Step 5.

 

Start the recording process by pressing the START 
button (position 2). The logging, paint and battery 
leds (position 4, 5 and 6) will blink once. The logging 
LED will blink blue at the specified logging interval.

 

 

Step 6

 

Place the CurveX 4 logger inside the insulation box.

 

Make sure the insulation

 

system is set up properly, as specified in the manual 
of the insulation box chosen. The cables are guided 
through the cable outlet and the cover is closed 
securely, using the gasket to ensure optimal heat 
protection.

 

 

Step 7

 

Send the system through the oven and then remove 
the instrument from the box as

 

soon as possible. The paint LED (position 5) blinks 
green after a full cure or red after a partial cure. The 
cure specification is set in the software, see the Ideal 
Finish Analysis help for more details (logger menu 

 

user settings).

 

 

Step 8

 

Stop the recording process by pressing the STOP 
button (position 3).

 

 

Step 9

 

Download the results with Ideal Finish Analysis:

 

 

Choose download all data to PC… in the Logger 

menu of Ideal Finish Analysis and follow the 
wizard.

 

Summary of Contents for CurveX4

Page 1: ...CurveX4 CX4005 CX4010 IMPORTANT Before taking this instrument in use we this manual carefully strongly advise you to read Operating Instructions V1 0 0522...

Page 2: ......

Page 3: ...with the operating instructions shall be excluded from thewarrantyaswellasmechanicalinterferenceby the customer or by third parties with CurveX 4 and relatedequipmentwithoutIndustrialPhysics written...

Page 4: ...e and Protective Measureswhennot inuse 4 Getting Started 8 4 1 Quick start 4 2 Memory batches 4 3 CurveX 4 tasks 4 4 Downloading Data And Configuring The CurveX 4 4 5 Logger menu download all data to...

Page 5: ...in this manual should have been read and understood acknowledged as evidenced by their signature 1 4 Dangers This instrument has been designed and constructed in accordance with state of the art tech...

Page 6: ...ty and dangers on the instrument are in readablecondition In case of danger the instrument has to be switched off by means of the on off switch at the left back side of the instrument or by unplugging...

Page 7: ...ssibly photographed and reported to the relevant insurers or in the case of sales deliveredtocustomersworks tothesupplier 3 4 StorageandProtectiveMeasureswhennot in use The instrument must be stored i...

Page 8: ...robes to the input points at position 7 of the CurveX 4 Step 3 Attach the sensors to the object to be measured Step 4 Press and hold down the Power button position 1 for one second to switch on the Cu...

Page 9: ...aint LED Red partial cure Orange active Green full cure 6 Battery LED Shows battery status when powered on Red 25 Orange 25 to 75 Green 75 7 Thermocouple connectors 1 4 Connect up to four probes 8 USB...

Page 10: ...play You will have to use Ideal Finish Analysis to download data and change the settings of the logger The license key and memory stick with the Ideal Finish Analysis software are part of the CurveX 4...

Page 11: ...Set the paint types in yourlogger Set the cure specifications for the report Configure the properties when logging data in Real Time Enter a description of the object you are measuring and enter a de...

Page 12: ...ure 2 Thermocouple K Connector Usage The position of the sensors on the object can be stored inIdealFinishAnalysis Thepositionwillthen also be available in thereports Remark Make sure that the probe c...

Page 13: ...is part in order to align the sensor element with the surface to ensure good contact Take the clamp between your thumb and forefinger Check which jaw has the sensor element and place the probe at the...

Page 14: ...heetforthe maximum time and temperatureuse 6 1 Preparing for measurement Step1 Makesurethe box is at room temperature Step 2 Place the CurveX 4 data logger with the probesalreadyconnected insidethebox...

Page 15: ...erve the elasticity of the rubber gasket 6 3 Precautions Batteries in carry on baggage aircraft cabin The battery employed in our CurveX 4 is a standard set of 2 AA batteries The batteries employed in...

Page 16: ...y 32 MB required hard Disk space 128 MB After300cycles charge from 0 to100 the batterywillretain acapacity of 80 Ingeneralsuchbatteryneedreplacementafterapprox 2yearscontinuoususe 7 1 CurveX 4 package...

Page 17: ...graphic is thus an absolute worst case scenario The green zone shows the worst case performance at standard environmental conditions Theorange part is its full designed temperature use Meaning the max...

Page 18: ...roduct whether inthis manual or other wise is correct we haveno control over either the quality or condition of the product orthemanyfactors affecting the use and application of the product Therefore...

Page 19: ...19...

Page 20: ...20 www industrialphysics com...

Reviews: