background image

 
 
 
 
 

 

 

 

automatic-film-applicator-ab3120-m44 - Page 4 of 23 

 

 

4 |  

1

 

GENERAL 

1.1

 

 Importance of operating manual  

This manual is written in order to become familiar with all the functions and possible applications of the 
instrument. It contains important instructions about how to use the instrument safely and economically; 
according to the purpose designated. Following these instructions is not only essential to avoid risks. It also 
reduces repair costs and down-time and increases the products reliability and service-life.  
Anyone who works with the instrument should follow the instructions in this manual, particularly the safety 
related instructions. Additionally local rules and regulations relating to environmental safety and accident 
prevention should be observed. 

 

1.2

 

User-responsibility  

The user should 

a)

   Only allow persons to work with the instrument who are familiar with the general instructions on how to 

work safely and to prevent accidents. The use of the instrument should have been instructed duly. The 
safety chapter and the warnings in this manual should have been read and understood; acknowledged as 
evidenced by their signature. 

b)

   Regularly check the safety-awareness of personnel at work.  

 

1.3

 

Responsibility of personnel  

Before commencing work anyone appointed to work with the instrument should pay attention to the general 
regulations relating to working safety and accident prevention. The safety chapter and the warnings in this 
manual should have been read and understood; acknowledged as evidenced by their signature. 

 

1.4

 

Dangers  

This instrument has been designed and constructed in accordance with state-of-the-art technology and the 
acknowledged safety regulations. Nevertheless, working with the instrument may cause danger to the life and 
health of the operator or to others, or damage to the instrument or other property. Therefore the instrument 
should only be used for its designated purpose, and in a perfect technical condition. Any defect that could have 
a negative effect on safety should be repaired immediately. 
 

1.5

 

Designated purpose  

The TQC Film applicator is exclusively designed to apply films of paint and coatings on test panels and test 
charts and as from models launched after Apr. 2013 for testing of the drying time of coatings on previous 
described substrates. 
Other applications constitute improper use. TQC will not be held liable for damage resulting from improper use.  
Designated purpose also includes properly observing all instructions in the operation manual, and adherence to 
inspection and maintenance schedules.  
 

1.6

 

Copyright  

The copyright of this operating manual remains with TQC. 
This operating manual is intended solely for the user and his personnel. Its instructions and guidelines may not 
be duplicated, circulated or otherwise passed on to others, neither fully, nor partly. Infringement of these 
restrictions may lead to legal action may be taken if this restrictions are infringed upon.  
 

1.7

 

Manufacturer's/Supplier's address  

TQC - Molenbaan 19, 2908 LL Capelle aan den IJssel - The Netherlands,  
T +31(0)10 7900100, F +31(0)10 7900129 

Summary of Contents for AB3120

Page 1: ...RECORDER AB3120 AB3220 AB3320 AB3400 AB3125 AB3225 AB3325 AB3405 IMPORTANT Before operating this instrument we strongly advise you to read this manual carefully Operating Instructions for the AUTOMAT...

Page 2: ...TRUCTIONS 5 2 1 Meaning of Symbols 5 2 2 Availability of Safety Information 5 2 3 Training of Personnel 5 2 4 Dangers from Electrical Energy 5 2 5 Points of Special Danger 5 2 6 Care Maintenance Repai...

Page 3: ...s Panel 12 7 4 Vacuum Suction Plate 12 7 5 Double Channel Vacuum 12 7 6 Film Applicators Tools 12 7 7 Test Charts 13 7 8 Drying Time Recorder 13 7 9 Multi Tool Platform 13 8 MENU DISPLAY INFORMATION A...

Page 4: ...this manual should have been read and understood acknowledged as evidenced by their signature 1 4 Dangers This instrument has been designed and constructed in accordance with state of the art technolo...

Page 5: ...the instrument are in readable condition In case of danger the instrument has to be switched off by means of the emergency button on the front of the instrument Then eliminate danger 2 3 Training of P...

Page 6: ...lier Parts used from other sources aren t guaranteed to take the loading and meet the safety requirements 2 8 Cleaning of the Instrument and Disposal of Materials When in use it is not always possible...

Page 7: ...uum bed C built in vacuum pump and combined attachment assembly for standard block applicators and wire bar coaters AB3400 TQC motorised automatic film applicator 230V with heated perforated vacuum be...

Page 8: ...d perforated vacuum bed Minimum temperature Ambient 5 C Maximum temperature Ambient 100 C Absolute max 140 C Resolution of set temperature 1 C Resolution of readout temperature 0 1 C Temperature contr...

Page 9: ...ecified on the indication label corresponds to the local supply voltage If it does not the instrument must not be connected under any circumstances Contact your local supplier or TQC for full specific...

Page 10: ...mation 2 Jog Shuttle 3 Emergency button 4 Acoustic alarm Buzzer 5 Levelling supports 6 Glass plate or Vacuum table 7 Automated clamping device for test charts 8 Mains connection 9 Spiral applicator we...

Page 11: ...11 7 INSTRUMENT COMPONENT ASSEMBLY 7 1 Overview 1 Housing 2 Splash shield 3 Tool Carrier 4 Hight Adjustment fixation 5 Spiral bar weight 6 Spiral bar weight guide pins 7 Spiral bar weight guide rings...

Page 12: ...the red supporting studs on the O rings The vacuum plate serves to hold thick foils charts and other papers for coating The area size under vacuum depends on the selection made in the menu of the ins...

Page 13: ...rder following steps need to be followed Remove the Spiral bar weight 5 from drawing chapter 7 by pulling out the Spiral bar weight guide pins 6 Take care that both Spiral bar weight guide pins 6 and...

Page 14: ...d as well as the length is shown in the top of the display Press the Jog shuttle knob to confirm or turn it to go back in the menu and make changes Place the draw down chart on the glass plate or vacu...

Page 15: ...e length is automatically set in accordance with the start end points in millimetres After made changes rotate to BACK and press the Jog shuttle knob VACUUM will only switch dependent on the length of...

Page 16: ...by selecting back NOTE The vacuum control is only available for models equipped with vacuum RUN SETUP LENGTH The length of a drying time test can be set here The operation is the same as that for sett...

Page 17: ...et knob on the front To set the temperature first turn the machine on and then set the temperature The test bed will warm up to the set temperature 8 4 Warning signals Due to circumstances the display...

Page 18: ...er use Clean the instrument using a soft dry cloth Never clean the instrument by any mechanical means such as a wire brush or abrasive paper This may cause just like the use of aggressive cleaning age...

Page 19: ...advice we give about the product whether in this manual or otherwise is correct we have no control over either the quality or condition of the product or the many factors affecting the use and applic...

Page 20: ...d differentiate in size Place the smallest of the rings A4 around the inner supports and the largest A3 around the outer supports as shown in below image detail A Gently lower down the vacuum plate B...

Page 21: ...automatic film applicator ab3120 m44 Page 21 of 23 21 ANNEX B POSITIONING OF SPIRAL BAR LIFTER...

Page 22: ...automatic film applicator ab3120 m44 Page 22 of 23 22 ANNEX C INSTALLATION OF THE LAMP...

Page 23: ...automatic film applicator ab3120 m44 Page 23 of 23 23...

Reviews: