8.2.10.4 Object 6504h
– Supported alarms
This object contains information on supported alarms by the measuring system.
Unsigned16
Bit
Function
Bit = 0
Bit = 1
0
Position error
No
Yes
1-15
Reserved for further use
8.2.10.5 Object 6505h
– Warnings
This object is not supported.
By read access the value is always "0".
8.2.10.6 Object 6506h
– Supported warnings
This object is not supported.
By read access the value is always "0".
8.2.10.7 Object 6507h
– Profile and software version
This object contains in the 1st 16 bits the profile version which is implemented in the
measuring system. It is combined to a revision number and an index.
e.g.:
Profile version:
1.40
Binary code:
0000 0001 0100 0000
Hexadecimal:
1
40
The 2nd 16 bits contain the software version which is implemented in the measuring
system. Only the last 4 digits are available.
e.g.:
Software version: 5022.01
Binary code:
0010 0010 0000 0001
Hexadecimal:
22
01
The complete software version is contained in object 100Ah, see page 154.