background image

Traceable

®

 is a registered trademark of Control Company   

©

2009  Control Company.  92-4330-00 Rev. 2 111809

 

 

 

 
SPECIFICATIONS

 

Ranges:

  

0.2 to 20.0 meters per second 

(air velocity) 

0.7 to 72.0 kilometers per hour 

 

40 to 3940 feet per minute 

 

0.5 to 44.7 miles per hour 

 

0.4 to 38.8 knots (nautical miles per hour) 

(temperature)  32.0 to 122.0°F (0.0 to 50.0°C) 

Resolution:

   For meters per second, resolution is 0.1 

(air velocity) 

For kilometers per hour, resolution is 0.1; 

 

For feet per minute, resolution is 1; 

 

For miles per hour, resolution is 0.1; 

 

For knots, resolution is 0.1. 

(temperature)  0.1°F (0.1°C) 

Accuracy:

 

± 1% full scale plus 1 digit of the reading 

(air velocity) 
(temperature)  0.8°C (1.5°F). 

Sampling Time:

 

approximately 0.8 seconds 

Data Output:

  Computer interface 

Case:

 

ABS plastic 

Fail safe:

 

Low battery indicator 

Power:

 

Six (6) 1.5V AAA alkaline batteries 

Accessories  
Supplied:

 

carrying case, telescope probe, batteries,  

 

Traceable

®

 Certificate, instructions. 

 

DESCRIPTION

 

3-1  LCD  
3-2  Power Button 
3-3  Hold Button: press to hold reading 
3-4  °C/°F button 
3-5  Record Button: press to record results. 
3-6  Recall Button: press to show minimum and maximum readings. 
3-7  Zero Button 
3-8  Unit button: press to select air velocity measurement 
3-9  Battery Compartment 
3-10  Computer Output Socket 
3-11  Probe Input Socket 
3-12  Sensor Head 
3-13  Probe Handle 
3-14  Probe Plug 
 

 
 
 

 
 
 
HOT WIRE ANEMOMETER OPERATION 

1. 

Insert the PROBE PLUG (3-14, fig. 1) into the socket on top of the unit labeled 
INPUT (3-11, fig. 1). 

2. 

Turn the meter on by switching the POWER BUTTON (3-2, fig. 1) to the right. 

3. 

Select the desired temperature by pressing the °C/°F BUTTON (3-4, fig. 1). 

4. 

Select the desired air velocity unit by pressing the UNIT BUTTON (3-8, fig. 1). 
Each press of the button will show a different unit of measurement on the 
display (m/s, km/h, ft/min, Knot, mile/h). 

5. 

Slide the sensor cover to the end of the SENSOR HEAD (3-12 fig 1). See Figure 
2. This allows the sensor head to be protected from the environment for a more 
accurate zero reading. 

 

Figure 2 

6. 

Push the ZERO BUTTON (3-7, fig. 1). The display 
should read 0,0. 

7. 

Slide the sensor cover towards the probe handle to 
expose the sensor head. 

Caution!! Do not allow 

fingers, tools or any other object to come in 
contact with the air velocity sensor. Any contact 
may permanently damage the sensor and the meter will not work properly.

 

8. 

Adjust the TELESCOPE PROBE (3-13, fig. 1) to the desired length. 

9. 

Note the mark located at the top of the SENSOR HEAD. This mark should be 
pointed towards the source of wind being measured. When this mark is placed 
towards the wind source, the upper display will show the air velocity value. The 
lower display shows the temperature. 

10.  To hold a measurement on the display, press and hold the HOLD BUTTON (3-3, 

fig. 1) while a measurement is being taken. The LCD display will show 

DH

 in the 

upper left portion of the display to indicate that the value is a “held” value. To 
cancel the data hold feature, simply press the HOLD BUTTON a second time. 

11.  To record a measurement, press and hold the RECORD BUTTON (3-5, fig. 1) 

while a measurement is being taken to hold the displayed value. The LCD 
display will show 

REC

 in the lower left portion of the display to indicate that the 

value is being recorded. To deactivate the record function, press the RECORD 
BUTTON again. 

MEMORY RECALL OPERATION

 

Use the RECALL BUTTON (3-6, fig. 1) to recall the last minimum and maximum 
readings. When the 

REC

 symbol appears on the display, press the RECALL BUTTON 

(3-6, fig 1) once. The maximum recorded value will be displayed. II letters “Max” will 
also appear on indicating that this is the maximum or highest reading. A second press 
of the RECALL BUTTON will display the minimum reading. The letters “Min” will also 
appear indicating that this is the minimum or lower reading. Note: The Data Record 
function must be in use to utilize the Memory recall features. Once the RECORD 
BUTTON has been pressed a second time to deactivate the data record function, the 
minimum and maximum values are no longer stored. 

COMPUTER INTERFACE 

This unit features computer output. A COMPUTER OUTPUT SOCKET (3-10, fig. 1) is 
located on the top of the unit. To utilize this feature, connect the unit to a PC with the 
Data Acquisition Accessory. (see Accessories section). 

LOW BATTERY 

If the letters “LBT” appear on the left corner of the display, ti indicates the batteries are 
low and need to be replaced. To replace the batteries, slide the battery cover located 
on the back of the unit away from the unit. Remove the old batteries and replace them 
with six new 1.5V AAA alkaline batteries. Use alkaline batteries, NOT regular or heavy 
duty batteries. Place the batteries in the proper direction as indicated by the diagram in 
the battery compartment. Incorrectly installed batteries may damage electronics. 
Replace battery cover. 

ALL OPERATIONAL DIFFICULTIES 

If this anemometer does not function properly for any reason, please replace the 
batteries with six new 1.5V AAA alkaline batteries (see Low Battery section, above). 
Low battery power can occasionally cause any number of “apparent” operational 
difficulties. Replacing the batteries with new fresh batteries will solve most difficulties. 

 

TRACEABLE

®

 HOT-WIRE ANEMOMETER/THERMOMETER INSTRUCTIONS  

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Reviews: