F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.
"FS" = Full scale in primary units.
FS
FS + 2%
√
FS + 1d
FS + 9d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
√
2
F9
Display
Divisions
Determines the desired weight increments. Value should be consis-
tent with legal requirements.
1
√
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
√
0.0
0.00 0.000
0.0000 00
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the
UNIT/SET
key one level begins the procedure.
Press
UNIT/SET
key to begin se-
quence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down with
the
UNIT/SET
key one level begins the procedure. NOTE – press and
hold the UNIT/SET for three seconds for effect.
Press
UNIT/SET
key to begin se-
quence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the
UNIT/SET
key one level begins the proce-
dure.
Press
UNIT/SET
key to begin se-
quence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the
UNIT/SET
key one level be-
gins the procedure.
Press
UNIT/SET
key to begin se-
quence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the
UNIT/SET
key one level
begins the procedure.
Press
UNIT/SET
key to begin se-
quence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Press the
UNIT/SET
key
twice to execute.
F24
Piece Count
Enable
Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count
0
√
1
Page 4-2
Neptune 4500