APPENDIX C: DETERMINING PROPER SPAN GAIN (F2)
C.1 SPAN GAIN OVERVIEW
The Span Gain parameter found in F2 of the Setup Menu is directly related to the ADC (Analog to
Digital Converter) integration time. This means that the lower the setting, the higher the number of
measurements per second. A span gain setting of
25
produces about 25 to 30 measurements per
second, while a span gain of
200
produces only about 3 or 4 measurements per second.
There is really no wrong setting for span gain – except in two cases. Using a low setting for a high
resolution, low output system could yield instability. Using a high setting in a high output system
could yield non-linearity.
C.2 SETTING THE INITIAL VALUE FOR SPAN GAIN
1. Determine the number of desired external graduations and choose the corresponding value
listed in Table C-1 or Table C-1a under the number closest to your full-scale input range in
millivolts.
2. Enter the Setup Menu and save this number for the Span Gain parameter in F2.
3. Perform a system calibration. If the calibration proves unsuccessful, or you wish to view the
internal counts, proceed to the next set of instructions.
C.3 VIEWING THE INTERNAL COUNTS
1. Enter the zero calibration menu (F16) and follow steps 1 to 3,
but do not save the zero point.
2. After pressing
ZERO
to zero the offset, place the test weight(s) on the platform. The displayed
count is the internal count. If the count remains on zero, check your load cell connections.
3. At full scale, the displayed count should be a minimum of 2 times the desired external
graduations. However, for maximum stability, a ratio of 6:1 or higher is recommended.
4. If the displayed count is large enough, remove the test weight(s), re-zero the indicator if
necessary, and proceed with the calibration. If the displayed number is
not
large enough,
increase the Span Gain to the next highest choice in the Setup Menu and re-calibrate.
Neptune 4500