background image

 
 

 

7.3 General Scale Operation 

7.3.1Force gauge 

 
1.  Select the desired weighing unit by pressing the 

UNITS

 key. 

2.  If necessary, press the 

ZERO

 key to obtain a weight reading of zero. 

3.  Place the object to be weighed on the platter and allow the indicator to stabilize. When 

the weighing unit default kN, if the item weight exceeds 9d over the system’s weight 
capacity, it displays “□□□□□□”. When the weight unit is N/lbf, if the item weight doesn’t 
exceed  9d  over  the  system’s  weight  capacity,  but  the  display  value  exceed  LED’s 
maximum digital, it display “□□□□□□”. 

4.  Read the weight shown on the display. 
 

7.3.2 Zero & Clear 

 
1.  In general force gauge mode, the key is

 ZERO

 function key: Press the 

ZERO

 key to 

obtain a weight reading of zero. 

2.  In keeping peak value mode, the key is 

CLEAR

 function key: Press the 

CLEAR 

key to 

clear  the  peak  value  which  is  shown  in  the  display, prepare  to  the  next  peak  value 
keeping operation. 

 

7.3.3 Keeping Peak Value 

 
1.  Press the 

PEAK

 key to enter keeping peak value mode. The item’s weight is from 2% 

to  100%  of  full  capacity,  the  indicator  can  record  the  maximum  value  in  real  time. 
When the weight is less 2% of full capacity, the function is unable. 

2.  If necessary, press 

CLEAR

 key to clear the present peak value shown in the display. 

3.  Press 

PEAK

 key again, exit the keeping peak value mode. 

4.  In  calibration  mode,  press 

PEAK

  key  to  enter  the  present  channel  to  set  the 

parameters. 

 

7.3.4 Channel Change 

 
1.  Press 

CHANNEL

  (        )  key  to  change  the  present  channel  circulating.  The  default 

channel is channel 0. When pressing the CHANNEL (        ) key, the system will automatic 
delay 2—5 seconds, then enter the next channel. TI-1500B has four absolute channels for 
custom using in the different load cell standard sources.   
 

Note: In general operate mode, just using CHANNEL key to change the channels. 
It’s different from using in calibration mode. 

 

7.3.5 Print 

 
1. This key is used to print weight information by serial printer and send weight information 
out to the PC by serial port.   

Summary of Contents for TI-1500B

Page 1: ...TI 1500B Force Gauge Digital Indicator Setup Operation Manual V A2...

Page 2: ...Descriptions 7 Chapter 6 Calibration 7 6 1 Calibration Overview 7 6 2 Zero Calibration F16 8 6 3 Span Calibration F17 8 6 4 View Calibration Values F18 9 6 5 Key in Zero Calibration Value F19 9 6 6 Ke...

Page 3: ...e the manual in a safe and convenient place so it will be available if you have questions concerning the operation of the scale If you are an installer the indicator s installation and wiring instruct...

Page 4: ...e into LOADCELL Port on the rear panel of the indicator Show in figure 2 2 TI 1500B Rear Panel Pin No Pin Name 1 8 Excitation 3 10 Excitation 5 12 Signal 7 14 Signal Figure 2 2 Pin assignments for the...

Page 5: ...ta channel Not need calibrate again In every channel the system has two menus The Setup Menu F Menu which configures the indicator to your force gauge s platform and the User Menu A Menu which configu...

Page 6: ...ss the PEAK SET key To exit without saving press the UNITS key to return to the current F heading 5 Repeat steps 1 through 4 until the Setup Menu is programmed Figure 3 1 Setup Menu Key Assignments 3...

Page 7: ...HANNEL key or PRINT key to move through the selection field 4 To save a new selection press the PEAK SET key To exit without saving press the UNITS key to return to the current A heading 5 Repeat step...

Page 8: ...tion is detected by comparing the present display update with the previous one Maximum value varies depending on local regulations 1d 3d 5d 10d F6 Digital Filter Averages weight readings to produce hi...

Page 9: ...a checkmark Name Code Description Code Value A1 Baud Rate Selects the baud rate for data transmission through the serial port 300 600 1200 2400 4800 9600 A2 Data Bits and Parity Selects the number of...

Page 10: ...tarily then revert back up to F16 At this time proceed to the F17 span calibration to complete indicator calibration 6 3 Span Calibration F17 1 While in the Setup mode scroll to F17 then scroll down o...

Page 11: ...F18 Note The values displayed in this procedure are valid only after a successful calibration has been performed using F16 and F17 1 While in the Setup mode scroll to F18 then scroll down once using t...

Page 12: ...or will prompt you to enter the information in the follow table 2 If the value shown is correct press PEAK SET key to save the value then move to the next value Otherwise use the four directional keys...

Page 13: ...he display detail of the LED LED Annunciator Meaning ZERO This light is active when ever the displayed weight is within zero range PEAK Keeping the peak value When the weight is less than 2 of full ca...

Page 14: ...eeping operation 7 3 3 Keeping Peak Value 1 Press the PEAK key to enter keeping peak value mode The item s weight is from 2 to 100 of full capacity the indicator can record the maximum value in real t...

Page 15: ...Full Scale input Signal 30mV including dead load Minimum Sensitivity Non H 44 0 4 V grad Minimum Sensitivity H 44 1 0 V grad Input Impedance 30M typical Calibration Method Software Calibration with l...

Page 16: ...activated by pressing the PRINT key on the indicator s front panel Figure B 1 shows a suggested cable diagram for interface to a PC Figure B 2 shows the serial data format for the Demand Mode Figure B...

Page 17: ...lue is larger than full scale capacity Use a smaller test weight or check keyed in value ERR1 Span Calibration Mode F17 Keyed in weight value is less than 1 of full scale capacity Use a larger test we...

Page 18: ...75 Deerfield Parkway Buffalo Grove IL 60089 U S A www transcell com Asia Pacific headquarters Tel 886 2 8919 1012 Fax 886 2 8919 1011 E mail info apac transcell com Add 7F No 129 Ln 235 Baoqiao Rd Xin...

Reviews: