Chapter 1: Introduction To The TI-1500B Digital Indicator 

         
          TI-1500B  indicator  is  always  used  in  many  industrial  fields,  such  as  measuring 
martial  performance,  measuring  peak  value,  checking  resistance  strain  gauge  load  cell 
and  so  on.  It  is  industrial  class  digital  indicator.  This  mode  operates  identically,  can 
readout up to 100,000 display divisions. It has 10/80Hz (Option) A/D converting frequent, 
four  absolute  channels.  All  setup  parameters  may  be  entered  via  the  front  panel  keys, 
including calibration. 
 
          If  your  Model  TI-1500B  digital  indicator  is  part  of  a  complete  force  gauge  or  has 
been installed for you, you may skip to Chapter 7 for operating instructions. Prior to using 
the indicator, please read this chapter carefully and completely. Store the manual in a safe 
and convenient place so it will be available if you have questions concerning the operation 
of the scale. 
 

If you are an installer, the indicator’s installation and wiring instructions are found in 

Chapter 2. The indicator contains two main setup menus: The Setup ( “F” ) menu, which 
configures the indicator to your force gauge’s platform and the User ( “A” ) menu, which 
configures the serial communication port and enables some user options. Chapter 3 gives 
an  overview  and  explains  how  to  use  the  five  front  panel  keys  to  maneuver  and  save 
settings  in  both  menus.  Chapters  4  and  5  explain  the  Setup  and  User  Menu  options, 
respectively. Chapter 6 covers system calibration. Prior to installing the indicator, please 
read  this  manual  carefully  and  completely.  Store  the  manual  in  a  safe  and  convenient 
place so it will be available if you have questions concerning the setup and operation of 
the force gauge. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Figure 1-1: TI-1500B Front Panel 

Summary of Contents for TI-1500B

Page 1: ...TI 1500B Force Gauge Digital Indicator Setup Operation Manual V A2...

Page 2: ...Descriptions 7 Chapter 6 Calibration 7 6 1 Calibration Overview 7 6 2 Zero Calibration F16 8 6 3 Span Calibration F17 8 6 4 View Calibration Values F18 9 6 5 Key in Zero Calibration Value F19 9 6 6 Ke...

Page 3: ...e the manual in a safe and convenient place so it will be available if you have questions concerning the operation of the scale If you are an installer the indicator s installation and wiring instruct...

Page 4: ...e into LOADCELL Port on the rear panel of the indicator Show in figure 2 2 TI 1500B Rear Panel Pin No Pin Name 1 8 Excitation 3 10 Excitation 5 12 Signal 7 14 Signal Figure 2 2 Pin assignments for the...

Page 5: ...ta channel Not need calibrate again In every channel the system has two menus The Setup Menu F Menu which configures the indicator to your force gauge s platform and the User Menu A Menu which configu...

Page 6: ...ss the PEAK SET key To exit without saving press the UNITS key to return to the current F heading 5 Repeat steps 1 through 4 until the Setup Menu is programmed Figure 3 1 Setup Menu Key Assignments 3...

Page 7: ...HANNEL key or PRINT key to move through the selection field 4 To save a new selection press the PEAK SET key To exit without saving press the UNITS key to return to the current A heading 5 Repeat step...

Page 8: ...tion is detected by comparing the present display update with the previous one Maximum value varies depending on local regulations 1d 3d 5d 10d F6 Digital Filter Averages weight readings to produce hi...

Page 9: ...a checkmark Name Code Description Code Value A1 Baud Rate Selects the baud rate for data transmission through the serial port 300 600 1200 2400 4800 9600 A2 Data Bits and Parity Selects the number of...

Page 10: ...tarily then revert back up to F16 At this time proceed to the F17 span calibration to complete indicator calibration 6 3 Span Calibration F17 1 While in the Setup mode scroll to F17 then scroll down o...

Page 11: ...F18 Note The values displayed in this procedure are valid only after a successful calibration has been performed using F16 and F17 1 While in the Setup mode scroll to F18 then scroll down once using t...

Page 12: ...or will prompt you to enter the information in the follow table 2 If the value shown is correct press PEAK SET key to save the value then move to the next value Otherwise use the four directional keys...

Page 13: ...he display detail of the LED LED Annunciator Meaning ZERO This light is active when ever the displayed weight is within zero range PEAK Keeping the peak value When the weight is less than 2 of full ca...

Page 14: ...eeping operation 7 3 3 Keeping Peak Value 1 Press the PEAK key to enter keeping peak value mode The item s weight is from 2 to 100 of full capacity the indicator can record the maximum value in real t...

Page 15: ...Full Scale input Signal 30mV including dead load Minimum Sensitivity Non H 44 0 4 V grad Minimum Sensitivity H 44 1 0 V grad Input Impedance 30M typical Calibration Method Software Calibration with l...

Page 16: ...activated by pressing the PRINT key on the indicator s front panel Figure B 1 shows a suggested cable diagram for interface to a PC Figure B 2 shows the serial data format for the Demand Mode Figure B...

Page 17: ...lue is larger than full scale capacity Use a smaller test weight or check keyed in value ERR1 Span Calibration Mode F17 Keyed in weight value is less than 1 of full scale capacity Use a larger test we...

Page 18: ...75 Deerfield Parkway Buffalo Grove IL 60089 U S A www transcell com Asia Pacific headquarters Tel 886 2 8919 1012 Fax 886 2 8919 1011 E mail info apac transcell com Add 7F No 129 Ln 235 Baoqiao Rd Xin...

Reviews: