2
Testing and Calibration
Calibrating a Wrist Strap Tester
For calibration of a wrist strap tester, turn the knob to
"
680K Ω LOW
” setting on the
Upper
part of dial marked
"Wrist strap”
.
1. Make sure that the test leads are connected to the cor-
rect color coded jacks. Then connect the opposite end
of one test leads to
WRIST STRAP
jack of tester
making sure to touch the metal side wall of the port.
2. Place the other probe onto the touch plate of the
tester, making sure that the probe is placed in the
middle of the touch plate for optimal reading.
Be sure to hold the cord at an insulated point, so that the
resistance value is not affected by the body.
Power Source
Assembly
The CAL1000 requires little
assembly. Locate the colored
jacks on the front of unit and
simply plug the
RED
test lead
into the
RED
jack and the
BLACK
test lead into the
BLACK
jack. You are now
ready to begin calibration.
Figure 1:CAL 1000
and test leads