Testing and Calibration

 

Calibrating a Wrist Strap Tester 
 

For calibration of a wrist strap tester, turn the knob to 
"

680K Ω LOW

” setting on the 

Upper

 part of dial marked 

"Wrist strap”

 
1.  Make sure that the test leads are connected to the cor-

rect color coded jacks. Then connect the opposite end 
of one test leads to 

WRIST STRAP

 jack of tester 

making sure to touch the metal side wall of the port. 

2.  Place the other probe onto the touch plate of the 

tester, making sure that the probe is placed in the 
middle of the touch plate for optimal reading.  

 
Be sure to hold the cord at an insulated point, so that the 
resistance value is not affected by the body.

 

Power Source

 

Assembly 

 

The CAL1000 requires little 
assembly.  Locate the colored 
jacks on the front of unit and 
simply plug the 

RED

 test lead 

into the 

RED

 jack and the 

BLACK

 test lead into the 

BLACK 

jack. You are now 

ready to begin calibration. 

Figure 1:CAL 1000  

and test leads 

Summary of Contents for Ohm Metrics CAL1000

Page 1: ...truction Manual Combo Tester Verification Device Model CAL1000 T R A N S F O R M I N G T E C H N O L O G I E S L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L Ohm Metrics ...

Page 2: ...ription CAL 1000 1 OHM Metrics 1 2 Installation Power Source 2 Assembly 2 3 Testing and Calibration Wrist Strap Calibration 2 Footwear Calibration 3 4 Parts Included List 4 5 Specifications 5 6 Service and Warranty 6 ...

Page 3: ...e Passive No power source required NIST traceable for in house calibration Wrist strap test range 750 kilohm 10 megohm Footwear test range 750 kilohm 100 megohm Versatile works with most tester brands Ohm Metrics test and measurement products from Transforming Technologies are useful reliable tools for characterizing and identifying the electrical resistance of materials and the performance of per...

Page 4: ...h the metal side wall of the port 2 Place the other probe onto the touch plate of the tester making sure that the probe is placed in the middle of the touch plate for optimal reading Be sure to hold the cord at an insulated point so that the resistance value is not affected by the body Power Source Assembly The CAL1000 requires little assembly Locate the colored jacks on the front of unit and simp...

Page 5: ...t strap tester should indicate a GREEN LED light PASS If the tester does not meet calibration specifications you should contact the supplier for adjustment information Calibrating a Footwear Tester For calibration of a footwear tester turn the knob to 680K Ω LOW setting on the LOWER part of dial marked Footwear Calibration test procedures 1 Make sure that the test leads are connected to the cor re...

Page 6: ...D LED light LO 4 Repeat this procedure for the other footwear knob set tings In each case the indication on the instrument should cor respond to the knob setting For example set the calibra tion unit to the PASS position repeat steps 1 3 and the footwear tester should indicate a GREEN LED light PASS Be sure to hold the cord at an insulated point so that the resistance value is not affected by the ...

Page 7: ... 7 x 3 7 x 1 3 Weight 0 41lbs Product Number Cal 1000 Calibration Unit Calibrates WST200 Wrist Strap Tester GTS600 Combo Tester GTS900 Combo Tester Calibrates Most Makes and Models of ESD Testers Resistance Range Wrist Strap LOW 0 68MΩ PASS 0 85 MΩ 3M 9 1MΩ HIGH 11MΩ Footwear LOW 0 68MKΩ PASS 820KΩ 38M 88MΩ HIGH 110MΩ ...

Page 8: ...es be liable for collat eral or consequential damages To receive service under warranty please contact Trans forming Technologies Technical Support About Transforming Technologies Since 1998 Transforming Technologies has helped elec tronic manufacturing facilities to protect their products and processes from the many serious problems associated with static electricity Transforming Technologies off...

Reviews: