Calibrating a Wrist Strap Tester Cont.
3. By pushing the probe on the touch plate you'll usually
activate the tester.
4. The tester should indicate
RED LED light "LO".
5. Repeat this procedure for the other wrist strap knob
settings.
In each case the indication on the instrument should cor-
respond to the knob setting. For example, set the calibra-
tion unit to the "PASS" position, repeat steps 1-3 and the
wrist strap tester should indicate a
GREEN LED light
“PASS”
.
If the tester does not meet calibration specifications, you
should contact the supplier for adjustment information.
Calibrating a Footwear Tester
For calibration of a footwear tester, turn the knob to
"
680K Ω LOW
” setting on the
LOWER
part of dial marked
"Footwear”
.
Calibration test procedures:
1. Make sure that the test leads are connected to the cor-
rect color coded jacks and then connect the opposite
end of one test leads to
Footwear
jack of tester.
Make sure that if the tester is a duel wrist strap/ footwear
tester, all adjustments are set to footwear.