Calibrating a Wrist Strap Tester Cont.  

 
3. By pushing the probe on the touch plate you'll usually 
activate the tester. 
4. The tester should indicate 

RED LED light "LO".

 

5. Repeat this procedure for the other wrist strap knob 
settings. 
  
In each case the indication on the instrument should cor-
respond to the knob setting. For example, set the calibra-
tion unit to the "PASS" position, repeat steps 1-3 and the 
wrist strap tester should indicate a 

GREEN LED light 

“PASS”

 
If the tester does not meet calibration specifications, you 
should contact the supplier for adjustment information. 

 
Calibrating a Footwear Tester 

 

For calibration of a footwear tester, turn the knob to 
"

680K Ω LOW

” setting on the 

LOWER

 part of dial marked 

"Footwear”

 
Calibration test procedures: 

 
1.  Make sure that the test leads are connected to the cor-

rect color coded jacks and then connect the opposite 
end of one test leads to 

Footwear

 jack of tester.  

 
Make sure that if the tester is a duel wrist strap/ footwear 
tester, all adjustments are set to footwear. 
 

Summary of Contents for Ohm Metrics CAL1000

Page 1: ...truction Manual Combo Tester Verification Device Model CAL1000 T R A N S F O R M I N G T E C H N O L O G I E S L L C O U T S T A N D I N G A L T E R N A T I V E S I N S T A T I C C O N T R O L Ohm Metrics ...

Page 2: ...ription CAL 1000 1 OHM Metrics 1 2 Installation Power Source 2 Assembly 2 3 Testing and Calibration Wrist Strap Calibration 2 Footwear Calibration 3 4 Parts Included List 4 5 Specifications 5 6 Service and Warranty 6 ...

Page 3: ...e Passive No power source required NIST traceable for in house calibration Wrist strap test range 750 kilohm 10 megohm Footwear test range 750 kilohm 100 megohm Versatile works with most tester brands Ohm Metrics test and measurement products from Transforming Technologies are useful reliable tools for characterizing and identifying the electrical resistance of materials and the performance of per...

Page 4: ...h the metal side wall of the port 2 Place the other probe onto the touch plate of the tester making sure that the probe is placed in the middle of the touch plate for optimal reading Be sure to hold the cord at an insulated point so that the resistance value is not affected by the body Power Source Assembly The CAL1000 requires little assembly Locate the colored jacks on the front of unit and simp...

Page 5: ...t strap tester should indicate a GREEN LED light PASS If the tester does not meet calibration specifications you should contact the supplier for adjustment information Calibrating a Footwear Tester For calibration of a footwear tester turn the knob to 680K Ω LOW setting on the LOWER part of dial marked Footwear Calibration test procedures 1 Make sure that the test leads are connected to the cor re...

Page 6: ...D LED light LO 4 Repeat this procedure for the other footwear knob set tings In each case the indication on the instrument should cor respond to the knob setting For example set the calibra tion unit to the PASS position repeat steps 1 3 and the footwear tester should indicate a GREEN LED light PASS Be sure to hold the cord at an insulated point so that the resistance value is not affected by the ...

Page 7: ... 7 x 3 7 x 1 3 Weight 0 41lbs Product Number Cal 1000 Calibration Unit Calibrates WST200 Wrist Strap Tester GTS600 Combo Tester GTS900 Combo Tester Calibrates Most Makes and Models of ESD Testers Resistance Range Wrist Strap LOW 0 68MΩ PASS 0 85 MΩ 3M 9 1MΩ HIGH 11MΩ Footwear LOW 0 68MKΩ PASS 820KΩ 38M 88MΩ HIGH 110MΩ ...

Page 8: ...es be liable for collat eral or consequential damages To receive service under warranty please contact Trans forming Technologies Technical Support About Transforming Technologies Since 1998 Transforming Technologies has helped elec tronic manufacturing facilities to protect their products and processes from the many serious problems associated with static electricity Transforming Technologies off...

Reviews: